ISO 12189 Standard for the Preclinical Evaluation of Posterior Spinal Stabilization Devices--II: A Parametric Comparative Study

Overview
Date 2015 Dec 18
PMID 26673809
Citations 11
Authors
Affiliations
Soon will be listed here.
Abstract

The International Standardization Organization (ISO) 12189 standard was recently introduced to preclinically evaluate and compare the mechanical properties of posterior stabilization devices. This scenario presents some new significant steps ahead over the vertebrectomy model recommended by American Society for Testing and Materials (ASTM) F1717 standard: the modular anterior support allows for describing a closer scenario to the effective clinical use as well as to test very flexible and dynamic posterior stabilization devices. Despite these significant advantages, ISO 12189 received little attention in the literature. Anatomical parameters depending on the spinal level were compared to the published data or original measurements on biplanar stereoradiography on 13 patients. Other mechanical variables, describing the test set-up design, were considered and all parameters were investigated using a numerical parametric finite element model. Stress values were calculated by also considering their worst-case combination. The standard set-up represents quite well the anatomy of an instrumented average thoracolumbar segment. The parametric comparative analysis demonstrates a significant (even beyond +350%) maximum increase in the stress on the device, compared to the standard currently in use. The anterior support stiffness plays the most detrimental effect (maximum stress increases up to 396%). The initial precompression step has an important role in determining the final stress values achieved at peak load (up to +76%). Moreover, when combining these two contributions, an even higher stress increase may be achieved (up to 473%). Despite the other anatomical parameters playing a secondary role, their worst-case combination demonstrates that a device could potentially undergo higher stresses than those reached according to standard suggestions (maximum increase of 22.4% at L1). Any user/designer should be aware of these effects when using ISO 12189 standard for the preclinical evaluation of posterior spinal stabilization devices.

Citing Articles

How to improve the mechanical safety of a novel spinal implant while saving costs and time.

Kienle A, Wilke H, Schroder C, Pietsch A JOR Spine. 2024; 7(4):e70026.

PMID: 39726899 PMC: 11669745. DOI: 10.1002/jsp2.70026.


Will the adjustment of insertional pedicle screw positions affect the risk of adjacent segment diseases biomechanically? An in-silico study.

Huang C, Liu Z, Wei Z, Fang Z, Xi Z, Cai P Front Surg. 2023; 9:1004642.

PMID: 36713678 PMC: 9877423. DOI: 10.3389/fsurg.2022.1004642.


The preoperative Hounsfield unit value at the position of the future screw insertion is a better predictor of screw loosening than other methods.

Li J, Zhang Z, Xie T, Song Z, Song Y, Zeng J Eur Radiol. 2022; 33(3):1526-1536.

PMID: 36241918 PMC: 9935714. DOI: 10.1007/s00330-022-09157-9.


Annulus Calibration Increases the Computational Accuracy of the Lumbar Finite Element Model.

Xu C, Xi Z, Fang Z, Zhang X, Wang N, Li J Global Spine J. 2022; 13(8):2310-2318.

PMID: 35293827 PMC: 10538312. DOI: 10.1177/21925682221081224.


Biomechanical Effects of Pedicle Screw Positioning on the Surgical Segment in Models After Oblique Lumbar Interbody Fusion: An in-silico Study.

Xu C, Huang C, Cai P, Fang Z, Wei Z, Liu F Int J Gen Med. 2022; 15:1047-1056.

PMID: 35140507 PMC: 8818966. DOI: 10.2147/IJGM.S352304.