Impact of Endobronchial Coiling on Segmental Bronchial Lumen in Treated and Untreated Lung Lobes: Correlation with Changes in Lung Volume, Clinical and Pulmonary Function Tests
Overview
Affiliations
Objectives: To assess the impact of endobronchial coiling on the segment bronchus cross-sectional area and volumes in patients with lung emphysema using quantitative chest-CT measurements.
Materials And Methods: Thirty patients (female = 15; median age = 65.36 years) received chest-CT before and after endobronchial coiling for lung volume reduction (LVR) between January 2010 and December 2014. Thin-slice (0.6 mm) non-enhanced image data sets were acquired both at end-inspiration and end-expiration using helical technique and 120 kV/100-150 mAs. Clinical response was defined as an increase in the walking distance (Six-minute walk test; 6MWT) after LVR-therapy. Additionally, pulmonary function test (PFT) measurements were used for clinical correlation.
Results: In the treated segmental bronchia, the cross-sectional lumen area showed significant reduction (p < 0.05) in inspiration and tendency towards enlargement in expiration (p > 0.05). In the ipsilateral lobes, the lumina showed no significant changes. In the contralateral lung, we found tendency towards increased cross-sectional area in inspiration (p = 0.06). Volumes of the treated segments correlated with the treated segmental bronchial lumina in expiration (r = 0.80, p < 0.001). Clinical correlation with changes in 6MWT/PFT showed a significant decrease of the inspiratory volume of the treated lobe in responders only.
Conclusion: Endobronchial coiling causes significant decrease in the cross-sectional area of treated segment bronchi in inspiration and a slight increase in expiration accompanied by a volume reduction.
Key Points: • Endobronchial coiling has indirect impact on cross-sectional area of treated segment bronchi • Volume changes of treated lobes correlate with changes in bronchial cross-sectional area • Coil-induced effects reflect their stabilizing and stiffening impact on lung parenchyma • Endobronchial coiling reduces bronchial collapsing compensating the loss of elasticity.
Roodenburg S, Klooster K, Hartman J, Kontogianni K, Brock J, Dittrich A Respiration. 2024; 104(1):58-68.
PMID: 39284292 PMC: 11731836. DOI: 10.1159/000541366.
Kayikci H, Cimen P, Katgi N Sisli Etfal Hastan Tıp Bul. 2024; 57(4):506-512.
PMID: 38268657 PMC: 10805047. DOI: 10.14744/SEMB.2023.06767.
Outcomes of bronchoscopic lung volume reduction coil treatment in patients with severe emphysema.
Ilgazli A, Elicora A, Sezer H, Kucuk S, Pasiyev F, Kusbeci Ciftci T Turk Gogus Kalp Damar Cerrahisi Derg. 2023; 31(4):561-567.
PMID: 38076002 PMC: 10704536. DOI: 10.5606/tgkdc.dergisi.2023.24534.
Endobronchial coils for emphysema: Dual mechanism of action on lobar residual volume reduction.
Hartman J, Shah P, Sciurba F, Herth F, Slebos D Respirology. 2020; 25(11):1160-1166.
PMID: 32267059 PMC: 7687244. DOI: 10.1111/resp.13816.
Herth F, Slebos D, Shah P, Hetzel M, Schmid-Bindert G, LaPrad A Respiration. 2019; 98(6):512-520.
PMID: 31743933 PMC: 6979432. DOI: 10.1159/000502100.