» Articles » PMID: 26336572

A Compact, Bone-Attached Robot for Mastoidectomy

Overview
Journal J Med Device
Date 2015 Sep 4
PMID 26336572
Citations 14
Authors
Affiliations
Soon will be listed here.
Abstract

Otologic surgery often involves a mastoidectomy, which is the removal of a portion of the mastoid region of the temporal bone, to safely access the middle and inner ear. The surgery is challenging because many critical structures are embedded within the bone, making them difficult to see and requiring a high level of accuracy with the surgical dissection instrument, a high-speed drill. We propose to automate the mastoidectomy portion of the surgery using a compact, bone-attached robot. The system described in this paper is a milling robot with four degrees-of-freedom (DOF) that is fixed to the patient during surgery using a rigid positioning frame screwed into the surface of the bone. The target volume to be removed is manually identified by the surgeon pre-operatively in a computed tomography (CT) scan and converted to a milling path for the robot. The surgeon attaches the robot to the patient in the operating room and monitors the procedure. Several design considerations are discussed in the paper as well as the proposed surgical workflow. The mean targeting error of the system in free space was measured to be 0.5 mm or less at vital structures. Four mastoidectomies were then performed in cadaveric temporal bones, and the error at the edges of the target volume was measured by registering a postoperative computed tomography (CT) to the pre-operative CT. The mean error along the border of the milled cavity was 0.38 mm, and all critical anatomical structures were preserved.

Citing Articles

Burr Hole Endoscopic Mastoidectomy: A Morphometric Cadaveric Study.

Lavergne P, Khoury T, Kang K, Sathe A, Kelly P, Evans J J Neurol Surg B Skull Base. 2024; 85(Suppl 2):e73-e79.

PMID: 39444778 PMC: 11495902. DOI: 10.1055/s-0043-1777674.


Endoscopic Endonasal Trans-Sphenoidal Minimally Invasive Pituitary Surgery with Image Guided Navigation System (Igns): Learning Experience of Ent Surgeon: First Author.

Ghosh D, Majhi S, Choudhary A, Samaddar S, Guha A, Kumar S Indian J Otolaryngol Head Neck Surg. 2024; 76(1):953-965.

PMID: 38440488 PMC: 10909064. DOI: 10.1007/s12070-023-04334-8.


Review of Enhanced Handheld Surgical Drills.

Usevitch D, Bronheim R, Cartagena-Reyes M, Ortiz-Babilonia C, Margalit A, Jain A Crit Rev Biomed Eng. 2023; 51(6):29-50.

PMID: 37824333 PMC: 10874117. DOI: 10.1615/CritRevBiomedEng.2023049106.


A Self-Configuring Deep Learning Network for Segmentation of Temporal Bone Anatomy in Cone-Beam CT Imaging.

Ding A, Lu A, Li Z, Sahu M, Galaiya D, Siewerdsen J Otolaryngol Head Neck Surg. 2023; 169(4):988-998.

PMID: 36883992 PMC: 11060418. DOI: 10.1002/ohn.317.


Robotic Milling of Electrode Lead Channels During Cochlear Implantation in an Model.

Hermann J, Mueller F, Schneider D, OToole Bom Braga G, Weber S Front Surg. 2021; 8:742147.

PMID: 34859039 PMC: 8631814. DOI: 10.3389/fsurg.2021.742147.


References
1.
Bell B, Stieger C, Gerber N, Arnold A, Nauer C, Hamacher V . A self-developed and constructed robot for minimally invasive cochlear implantation. Acta Otolaryngol. 2012; 132(4):355-60. DOI: 10.3109/00016489.2011.642813. View

2.
Balachandran R, Mitchell J, Dawant B, Fitzpatrick J . Accuracy evaluation of microTargeting Platforms for deep-brain stimulation using virtual targets. IEEE Trans Biomed Eng. 2009; 56(1):37-44. PMC: 4431767. DOI: 10.1109/TBME.2008.2002110. View

3.
Dillon N, Balachandran R, Dit Falisse A, Wanna G, Labadie R, Withrow T . Preliminary Testing of a Compact, Bone-Attached Robot for Otologic Surgery. Proc SPIE Int Soc Opt Eng. 2014; 9036:903614. PMC: 4249709. DOI: 10.1117/12.2043875. View

4.
Maes F, Collignon A, Vandermeulen D, Marchal G, Suetens P . Multimodality image registration by maximization of mutual information. IEEE Trans Med Imaging. 1997; 16(2):187-98. DOI: 10.1109/42.563664. View

5.
Noble J, Warren F, Labadie R, Dawant B . Automatic segmentation of the facial nerve and chorda tympani in CT images using spatially dependent feature values. Med Phys. 2009; 35(12):5375-84. PMC: 2673604. DOI: 10.1118/1.3005479. View