Does Fat Suppression Via Chemically Selective Saturation Affect R2*-MRI for Transfusional Iron Overload Assessment? A Clinical Evaluation at 1.5T and 3T

Overview
Journal Magn Reson Med
Publisher Wiley
Specialty Radiology
Date 2015 Aug 27
PMID 26308155
Citations 17
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: Fat suppression (FS) via chemically selective saturation (CHESS) eliminates fat-water oscillations in multiecho gradient echo (mGRE) R2*-MRI. However, for increasing R2* values as seen with increasing liver iron content (LIC), the water signal spectrally overlaps with the CHESS band, which may alter R2*. We investigated the effect of CHESS on R2* and developed a heuristic correction for the observed CHESS-induced R2* changes.

Methods: Eighty patients [female, n = 49; male, n = 31; mean age (± standard deviation), 18.3 ± 11.7 y] with iron overload were scanned with a non-FS and a CHESS-FS mGRE sequence at 1.5T and 3T. Mean liver R2* values were evaluated using three published fitting approaches. Measured and model-corrected R2* values were compared and statistically analyzed.

Results: At 1.5T, CHESS led to a systematic R2* reduction (P < 0.001 for all fitting algorithms) especially toward higher R2*. Our model described the observed changes well and reduced the CHESS-induced R2* bias after correction (linear regression slopes: 1.032/0.927/0.981). No CHESS-induced R2* reductions were found at 3T.

Conclusion: The CHESS-induced R2* bias at 1.5T needs to be considered when applying R2*-LIC biopsy calibrations for clinical LIC assessment, which were established without FS at 1.5T. The proposed model corrects the R2* bias and could therefore improve clinical iron overload assessment based on linear R2*-LIC calibrations. Magn Reson Med 76:591-601, 2016. © 2015 Wiley Periodicals, Inc.

Citing Articles

Quantification of Liver Iron Overload with MRI: Review and Guidelines from the ESGAR and SAR.

Reeder S, Yokoo T, Franca M, Hernando D, Alberich-Bayarri A, Alustiza J Radiology. 2023; 307(1):e221856.

PMID: 36809220 PMC: 10068892. DOI: 10.1148/radiol.221856.


Quantitative MRI of diffuse liver diseases: techniques and tissue-mimicking phantoms.

Tipirneni-Sajja A, Brasher S, Shrestha U, Johnson H, Morin C, Satapathy S MAGMA. 2022; 36(4):529-551.

PMID: 36515810 DOI: 10.1007/s10334-022-01053-z.


Evaluation of Hepatic Iron Overload Using a Contemporary 0.55 T MRI System.

Campbell-Washburn A, Mancini C, Conrey A, Edwards L, Shanbhag S, Wood J J Magn Reson Imaging. 2021; 55(6):1855-1863.

PMID: 34668604 PMC: 9018883. DOI: 10.1002/jmri.27950.


Hepatic Iron Quantification Using a Free-Breathing 3D Radial Gradient Echo Technique and Validation With a 2D Biopsy-Calibrated R Relaxometry Method.

Rohani S, Morin C, Zhong X, Kannengiesser S, Shrestha U, Goode C J Magn Reson Imaging. 2021; 55(5):1407-1416.

PMID: 34545639 PMC: 10424632. DOI: 10.1002/jmri.27921.


Quantification of pancreatic iron overload and fat infiltration and their correlation with glucose disturbance in pediatric thalassemia major patients.

Huang J, Shen J, Yang Q, Cheng Z, Chen X, Yu T Quant Imaging Med Surg. 2021; 11(2):665-675.

PMID: 33532266 PMC: 7779935. DOI: 10.21037/qims-20-292.


References
1.
Meloni A, Rienhoff Jr H, Jones A, Pepe A, Lombardi M, Wood J . The use of appropriate calibration curves corrects for systematic differences in liver R2* values measured using different software packages. Br J Haematol. 2013; 161(6):888-91. PMC: 3672338. DOI: 10.1111/bjh.12296. View

2.
Hernando D, Cook R, Diamond C, Reeder S . Magnetic susceptibility as a B0 field strength independent MRI biomarker of liver iron overload. Magn Reson Med. 2013; 70(3):648-56. PMC: 3883906. DOI: 10.1002/mrm.24848. View

3.
Robson M, Gatehouse P . Consequences of T2 relaxation during half-pulse slice selection for ultrashort TE imaging. Magn Reson Med. 2010; 64(2):610-5. DOI: 10.1002/mrm.22323. View

4.
Sanches-Rocha L, Serpa B, Figueiredo E, Hamerschlak N, Baroni R . Comparison between multi-echo T2* with and without fat saturation pulse for quantification of liver iron overload. Magn Reson Imaging. 2013; 31(10):1704-8. DOI: 10.1016/j.mri.2013.07.013. View

5.
Bydder M, Yokoo T, Hamilton G, Middleton M, Chavez A, Schwimmer J . Relaxation effects in the quantification of fat using gradient echo imaging. Magn Reson Imaging. 2007; 26(3):347-59. PMC: 2386876. DOI: 10.1016/j.mri.2007.08.012. View