Metal Artifact Reduction: Added Value of Rapid-Kilovoltage-Switching Dual-Energy CT in Relation to Single-Energy CT in a Piglet Animal Model

Overview
Specialties Oncology
Radiology
Date 2015 Aug 22
PMID 26295672
Citations 7
Authors
Affiliations
Soon will be listed here.
Abstract

Objective: The purpose of this article is to evaluate virtual monochromatic spectral imaging and metal artifact reduction software for reducing metal artifact and to compare it with conventional single-energy CT (SECT) in an animal model.

Materials And Methods: Postmortem juvenile (n = 5) and adult (n = 1) swine specimens were scanned with SECT followed by a dual-energy CT (DECT) pediatric protocol after the insertion of two rods into their paraspinal thoracolumbar regions. Virtual monochromatic spectral images were extrapolated from DECT images at five monoenergetic levels (64, 69, 75, 88, and 105 keV) with and without the use of metal artifact reduction software. Images were evaluated by a 5-point scoring system for the extent of metallic artifacts and image interpretability in soft-tissue and bone windows. The density in the most pronounced artifact was measured. CT dose index was recorded.

Results: In studies without metal artifact reduction software, higher energy reconstructions resulted in fewer artifacts and better image interpretability in both soft-tissue and bone windows (p < 0.0001). Artifact density decreased from -792 HU at 64 keV to -128 HU at 105 keV without the use of metal artifact reduction software. No difference was noted in attributes' scores or in artifact density in studies using metal artifact reduction software (p > 0.05). DECT studies showed lower scores compared with SECT with regard to all attributes. A new faint perimetallic hypodense halo was seen in all studies with metal artifact reduction software. The CT dose index of DECT was 1.18-3.56 times higher than that of SECT techniques.

Conclusion: DECT at all energy levels with metal artifact reduction software and higher energy extrapolations without metal artifact reduction software reduced metallic artifact and enhanced image interpretability compared with SECT. Radiation dose with DECT could be significantly higher than SECT.

Citing Articles

Use of virtual monoenergetic images for reduction of extensive dental implant associated artifacts in photon-counting detector CT.

Layer Y, Mesropyan N, Kupczyk P, Luetkens J, Isaak A, Dell T Sci Rep. 2024; 14(1):497.

PMID: 38177651 PMC: 10766624. DOI: 10.1038/s41598-023-50926-3.


Utility of an automatic adaptive iterative metal artifact reduction AiMAR algorithm in improving CT imaging of patients with hip prostheses evaluated for suspected bladder malignancy.

Mohammadinejad P, Khandelwal A, Inoue A, Takahashi H, Yalon M, Long Z Abdom Radiol (NY). 2022; 47(6):2158-2167.

PMID: 35320381 DOI: 10.1007/s00261-022-03475-8.


Reduction of microwave ablation needle related metallic artifacts using virtual monoenergetic images from dual-layer detector spectral CT in a rabbit model with VX2 tumor.

Wang G, Gao Q, Wang Z, Lu X, Yu S, Jin Z Sci Rep. 2021; 11(1):9295.

PMID: 33927327 PMC: 8085209. DOI: 10.1038/s41598-021-88853-w.


Metal artifact reduction techniques for single energy CT and dual-energy CT with various metal materials.

Kawahara D, Ozawa S, Yokomachi K, Higaki T, Shiinoki T, Saito A BJR Open. 2020; 1(1):20180045.

PMID: 33178930 PMC: 7592440. DOI: 10.1259/bjro.20180045.


Value of virtual monochromatic spectral images with metal artifact reduction algorithm in dual-energy computed tomography-guided microcoil localization of pulmonary nodules.

Liu Z, Zhang Z, Chen C, Hong N Medicine (Baltimore). 2018; 97(29):e11562.

PMID: 30024555 PMC: 6086503. DOI: 10.1097/MD.0000000000011562.