Selecting the Reference Image for Registration of CEST Series

Overview
Date 2015 Aug 14
PMID 26268435
Citations 34
Authors
Affiliations
Soon will be listed here.
Abstract

Background: To compare different reference images selected for registration among chemical exchange saturation transfer (CEST) series.

Materials And Methods: Five normal volunteers and eight brain tumor patients were studied on a 3 Tesla scanner. Image registration was performed by choosing each of the acquired CEST saturation or unsaturation dynamic images as the reference. CEST images at 3.5 ppm (amide proton transfer, APT) were computed for each motion-corrected data set after main magnetic field inhomogeneity correction. A uniformity index was defined to quantify the efficacy of image registration using different reference images. Joint histograms and the structural similarity index (SSIM) were used to analyze the intrinsic image similarity between various dynamic images.

Results: Image registration increased the average uniformity index by 18% if the 3.5 ppm saturated image was selected as the reference image. However, registering to the unsaturated dynamic image reduced the uniformity index by 13% on average. The joint histogram analysis showed that the saturated dynamic images were highly similar (SSIM = 0.89 ± 0.01), and were considerably different from the unsaturated dynamic image (SSIM = 0.58 ± 0.03).

Conclusion: The selection of the 3.5 ppm dynamic image as the reference image generated the highest uniformity index for APT imaging though other saturated images were equally suited as reference images.

Citing Articles

State-specific GluCEST alterations in insular subregions are associated with depression and plasma inflammatory biomarker levels in patients with inflammatory bowel disease.

Xu L, Lu J, Zhou M, Shi H, Zheng J, Cheng T Neuroimage Clin. 2024; 45:103713.

PMID: 39608228 PMC: 11636208. DOI: 10.1016/j.nicl.2024.103713.


Motion and magnetic field inhomogeneity correction techniques for chemical exchange saturation transfer (CEST) MRI: A contemporary review.

Simegn G, Sun P, Zhou J, Kim M, Reddy R, Zu Z NMR Biomed. 2024; 38(1):e5294.

PMID: 39532518 PMC: 11606773. DOI: 10.1002/nbm.5294.


Calibration-free whole-brain CEST imaging at 7T with parallel transmit pulse design for saturation homogeneity utilizing universal pulses (PUSHUP).

Volzke Y, Akbey S, Lowen D, Pracht E, Stirnberg R, Gras V Magn Reson Med. 2024; 93(2):630-642.

PMID: 39301770 PMC: 11604840. DOI: 10.1002/mrm.30305.


Multi-pool chemical exchange saturation transfer MRI in glioma grading, molecular subtyping and evaluating tumor proliferation.

Zhu H, Li Y, Ding Y, Liu Y, Shen N, Xie Y J Neurooncol. 2024; 169(2):287-297.

PMID: 38874844 DOI: 10.1007/s11060-024-04729-9.


Whole-cerebrum guanidino and amide CEST mapping at 3 T by a 3D stack-of-spirals gradient echo acquisition.

Wang K, Ju L, Song Y, Blair L, Xie K, Liu C Magn Reson Med. 2024; 92(4):1456-1470.

PMID: 38748853 PMC: 11262991. DOI: 10.1002/mrm.30134.


References
1.
Sakata A, Okada T, Yamamoto A, Kanagaki M, Fushimi Y, Okada T . Grading glial tumors with amide proton transfer MR imaging: different analytical approaches. J Neurooncol. 2015; 122(2):339-48. DOI: 10.1007/s11060-014-1715-8. View

2.
Hoffmann M, Carpenter T, Williams G, Sawiak S . A survey of patient motion in disorders of consciousness and optimization of its retrospective correction. Magn Reson Imaging. 2014; 33(3):346-50. DOI: 10.1016/j.mri.2014.11.004. View

3.
Smith S . Fast robust automated brain extraction. Hum Brain Mapp. 2002; 17(3):143-55. PMC: 6871816. DOI: 10.1002/hbm.10062. View

4.
Rohde G, Barnett A, Basser P, Marenco S, Pierpaoli C . Comprehensive approach for correction of motion and distortion in diffusion-weighted MRI. Magn Reson Med. 2004; 51(1):103-14. DOI: 10.1002/mrm.10677. View

5.
Wang Z, Bovik A, Sheikh H, Simoncelli E . Image quality assessment: from error visibility to structural similarity. IEEE Trans Image Process. 2004; 13(4):600-12. DOI: 10.1109/tip.2003.819861. View