» Articles » PMID: 26175571

Intervention Planning Using a Laser Navigation System for CT-Guided Interventions: A Phantom and Patient Study

Overview
Journal Korean J Radiol
Specialty Radiology
Date 2015 Jul 16
PMID 26175571
Citations 5
Authors
Affiliations
Soon will be listed here.
Abstract

Objective: To investigate the accuracy, efficiency and radiation dose of a novel laser navigation system (LNS) compared to those of free-handed punctures on computed tomography (CT).

Materials And Methods: Sixty punctures were performed using a phantom body to compare accuracy, timely effort, and radiation dose of the conventional free-handed procedure to those of the LNS-guided method. An additional 20 LNS-guided interventions were performed on another phantom to confirm accuracy. Ten patients subsequently underwent LNS-guided punctures.

Results: The phantom 1-LNS group showed a target point accuracy of 4.0 ± 2.7 mm (freehand, 6.3 ± 3.6 mm; p = 0.008), entrance point accuracy of 0.8 ± 0.6 mm (freehand, 6.1 ± 4.7 mm), needle angulation accuracy of 1.3 ± 0.9° (freehand, 3.4 ± 3.1°; p < 0.001), intervention time of 7.03 ± 5.18 minutes (freehand, 8.38 ± 4.09 minutes; p = 0.006), and 4.2 ± 3.6 CT images (freehand, 7.9 ± 5.1; p < 0.001). These results show significant improvement in 60 punctures compared to freehand. The phantom 2-LNS group showed a target point accuracy of 3.6 ± 2.5 mm, entrance point accuracy of 1.4 ± 2.0 mm, needle angulation accuracy of 1.0 ± 1.2°, intervention time of 1.44 ± 0.22 minutes, and 3.4 ± 1.7 CT images. The LNS group achieved target point accuracy of 5.0 ± 1.2 mm, entrance point accuracy of 2.0 ± 1.5 mm, needle angulation accuracy of 1.5 ± 0.3°, intervention time of 12.08 ± 3.07 minutes, and used 5.7 ± 1.6 CT-images for the first experience with patients.

Conclusion: Laser navigation system improved accuracy, duration of intervention, and radiation dose of CT-guided interventions.

Citing Articles

Navigation and Robotics in Interventional Oncology: Current Status and Future Roadmap.

Charalampopoulos G, Bale R, Filippiadis D, Odisio B, Wood B, Solbiati L Diagnostics (Basel). 2024; 14(1).

PMID: 38201407 PMC: 10795729. DOI: 10.3390/diagnostics14010098.


Scout-guided needle placement-a technical approach for dose reduction in CT-guided periradicular infiltration.

Nunninger M, Braun V, Ziegert M, Schwarz F, Hamm B, Scheel M Neuroradiology. 2019; 62(3):341-346.

PMID: 31838562 DOI: 10.1007/s00234-019-02339-1.


A radiopaque 3D printed, anthropomorphic phantom for simulation of CT-guided procedures.

Jahnke P, Schwarz F, Ziegert M, Almasi T, Abdelhadi O, Nunninger M Eur Radiol. 2018; 28(11):4818-4823.

PMID: 29789910 DOI: 10.1007/s00330-018-5481-4.


Infrared needle mapping to assist biopsy procedures and training.

Shar B, Leis J, Coucher J Healthc Technol Lett. 2018; 5(2):65-69.

PMID: 29750115 PMC: 5933369. DOI: 10.1049/htl.2017.0045.


Computed tomography-guided biopsies in children: accuracy, efficiency and dose usage.

Gruber-Rouh T, Thalhammer A, Klingebiel T, Nour-Eldin N, Vogl T, Eichler K Ital J Pediatr. 2017; 43(1):4.

PMID: 28057068 PMC: 5217564. DOI: 10.1186/s13052-016-0319-7.

References
1.
Nitta N, Takahashi M, Tanaka T, Takazakura R, Sakashita Y, Furukawa A . Laser-guided computed tomography puncture system: simulation experiments using artificial phantom lesions and preliminary clinical experience. Radiat Med. 2007; 25(4):187-93. DOI: 10.1007/s11604-006-0116-0. View

2.
Magnusson A, Radecka E, Lonnemark M, Raland H . Computed-tomography-guided punctures using a new guidance device. Acta Radiol. 2005; 46(5):505-9. DOI: 10.1080/02841850510021508. View

3.
Proschek D, Kafchitsas K, Rauschmann M, Kurth A, Vogl T, Geiger F . Reduction of radiation dose during facet joint injection using the new image guidance system SabreSource: a prospective study in 60 patients. Eur Spine J. 2008; 18(4):546-53. PMC: 2899461. DOI: 10.1007/s00586-008-0832-5. View

4.
Zangos S, Muller C, Mayer F, Naguib N, Nour-Eldin N, Hansmann M . [Retrospective 5-year analysis of MR-guided biopsies in a low-field MR system]. Rofo. 2009; 181(7):658-63. DOI: 10.1055/s-0028-1109349. View

5.
Becker H, Meissner O, Waggershauser T . [C-arm CT-guided 3D navigation of percutaneous interventions]. Radiologe. 2009; 49(9):852-5. DOI: 10.1007/s00117-009-1866-3. View