Intervention Planning Using a Laser Navigation System for CT-Guided Interventions: A Phantom and Patient Study
Overview
Authors
Affiliations
Objective: To investigate the accuracy, efficiency and radiation dose of a novel laser navigation system (LNS) compared to those of free-handed punctures on computed tomography (CT).
Materials And Methods: Sixty punctures were performed using a phantom body to compare accuracy, timely effort, and radiation dose of the conventional free-handed procedure to those of the LNS-guided method. An additional 20 LNS-guided interventions were performed on another phantom to confirm accuracy. Ten patients subsequently underwent LNS-guided punctures.
Results: The phantom 1-LNS group showed a target point accuracy of 4.0 ± 2.7 mm (freehand, 6.3 ± 3.6 mm; p = 0.008), entrance point accuracy of 0.8 ± 0.6 mm (freehand, 6.1 ± 4.7 mm), needle angulation accuracy of 1.3 ± 0.9° (freehand, 3.4 ± 3.1°; p < 0.001), intervention time of 7.03 ± 5.18 minutes (freehand, 8.38 ± 4.09 minutes; p = 0.006), and 4.2 ± 3.6 CT images (freehand, 7.9 ± 5.1; p < 0.001). These results show significant improvement in 60 punctures compared to freehand. The phantom 2-LNS group showed a target point accuracy of 3.6 ± 2.5 mm, entrance point accuracy of 1.4 ± 2.0 mm, needle angulation accuracy of 1.0 ± 1.2°, intervention time of 1.44 ± 0.22 minutes, and 3.4 ± 1.7 CT images. The LNS group achieved target point accuracy of 5.0 ± 1.2 mm, entrance point accuracy of 2.0 ± 1.5 mm, needle angulation accuracy of 1.5 ± 0.3°, intervention time of 12.08 ± 3.07 minutes, and used 5.7 ± 1.6 CT-images for the first experience with patients.
Conclusion: Laser navigation system improved accuracy, duration of intervention, and radiation dose of CT-guided interventions.
Navigation and Robotics in Interventional Oncology: Current Status and Future Roadmap.
Charalampopoulos G, Bale R, Filippiadis D, Odisio B, Wood B, Solbiati L Diagnostics (Basel). 2024; 14(1).
PMID: 38201407 PMC: 10795729. DOI: 10.3390/diagnostics14010098.
Nunninger M, Braun V, Ziegert M, Schwarz F, Hamm B, Scheel M Neuroradiology. 2019; 62(3):341-346.
PMID: 31838562 DOI: 10.1007/s00234-019-02339-1.
A radiopaque 3D printed, anthropomorphic phantom for simulation of CT-guided procedures.
Jahnke P, Schwarz F, Ziegert M, Almasi T, Abdelhadi O, Nunninger M Eur Radiol. 2018; 28(11):4818-4823.
PMID: 29789910 DOI: 10.1007/s00330-018-5481-4.
Infrared needle mapping to assist biopsy procedures and training.
Shar B, Leis J, Coucher J Healthc Technol Lett. 2018; 5(2):65-69.
PMID: 29750115 PMC: 5933369. DOI: 10.1049/htl.2017.0045.
Computed tomography-guided biopsies in children: accuracy, efficiency and dose usage.
Gruber-Rouh T, Thalhammer A, Klingebiel T, Nour-Eldin N, Vogl T, Eichler K Ital J Pediatr. 2017; 43(1):4.
PMID: 28057068 PMC: 5217564. DOI: 10.1186/s13052-016-0319-7.