» Articles » PMID: 19517340

[Retrospective 5-year Analysis of MR-guided Biopsies in a Low-field MR System]

Overview
Journal Rofo
Publisher Thieme
Date 2009 Jun 12
PMID 19517340
Citations 5
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: The purpose of this study was to evaluate the safety and clinical value of MR-guided biopsies in an open 0.2 T low-field system.

Materials And Methods: A total of 322 patients with suspicious lesions of different body regions were biopsied in a low-field MRI system (0.2 T, Concerto, Siemens). The procedures were guided using T 1-weighted Flash sequences (TR/TE = 100/9; 70 degrees). The lesions were repeatedly biopsied using the coaxial technique with a 15-gauge (diameter 2 mm) puncture needle. Complications and biopsy findings were analyzed retrospectively.

Results: In all cases the biopsy procedures were successfully performed with MR guidance. In 298 patients diagnosis was able to be confirmed on the basis of the probes. The clinical follow-up showed that in 24 patients the lesions were missed by MR-guided biopsy. From this a sensitivity of 86%, a specificity of 87% and an accuracy of 93% were calculated. In two patients major complications were observed (morbidity rate 0.6 %).

Conclusion: MR-guided biopsy can be performed safely and precisely in a low-field MR system and are a supplement to US or CT-guided biopsies.

Citing Articles

Computed tomography-guided biopsies in children: accuracy, efficiency and dose usage.

Gruber-Rouh T, Thalhammer A, Klingebiel T, Nour-Eldin N, Vogl T, Eichler K Ital J Pediatr. 2017; 43(1):4.

PMID: 28057068 PMC: 5217564. DOI: 10.1186/s13052-016-0319-7.


Navigated MRI-guided liver biopsies in a closed-bore scanner: experience in 52 patients.

Moche M, Heinig S, Garnov N, Fuchs J, Petersen T, Seider D Eur Radiol. 2015; 26(8):2462-70.

PMID: 26563349 PMC: 4927599. DOI: 10.1007/s00330-015-4097-1.


Intervention Planning Using a Laser Navigation System for CT-Guided Interventions: A Phantom and Patient Study.

Gruber-Rouh T, Lee C, Bolck J, Naguib N, Schulz B, Eichler K Korean J Radiol. 2015; 16(4):729-35.

PMID: 26175571 PMC: 4499536. DOI: 10.3348/kjr.2015.16.4.729.


Performing MR-guided biopsies in clinical routine: factors that influence accuracy and procedure time.

Hoffmann R, Thomas C, Rempp H, Schmidt D, Pereira P, Claussen C Eur Radiol. 2011; 22(3):663-71.

PMID: 21960160 DOI: 10.1007/s00330-011-2297-x.


Magnetic resonance-guided upper abdominal biopsies in a high-field wide-bore 3-T MRI system: feasibility, handling, and needle artefacts.

Kuhn J, Langner S, Hegenscheid K, Evert M, Kickhefel A, Hosten N Eur Radiol. 2010; 20(10):2414-21.

PMID: 20503050 DOI: 10.1007/s00330-010-1809-4.