Low-dose CT Screening Using Hybrid Iterative Reconstruction: Confidence Ratings of Diagnoses of Simulated Lesions Other Than Lung Cancer

Overview
Journal Br J Radiol
Specialty Radiology
Date 2015 Jul 9
PMID 26153902
Citations 1
Authors
Affiliations
Soon will be listed here.
Abstract

Objective: To evaluate the confidence ratings of diagnoses of simulated lesions other than lung cancer on low-dose screening CT with hybrid iterative reconstruction (IR).

Methods: Simulated lesions (emphysema, mediastinal masses and interstitial pneumonia) in a chest phantom were scanned by a 320-row area detector CT. The scans were performed by 64-row and 160-row helical scans at various dose levels and were reconstructed by filtered back projection (FBP) and IR. Emphysema, honeycombing and reticular opacity were visually scored on a four-point scale by six thoracic radiologists. The ground-glass opacity as a percentage of total lung volume (%GGO), CT value and contrast-to-noise ratio (CNR) of mediastinal masses were calculated. These scores and values were compared between FBP and IR. Wilcoxon's signed-rank test was used (p < 0.05). Interobserver agreements were evaluated by κ statistics.

Results: There were no significant differences in visual assessment. Interobserver agreement was almost perfect. CT values were almost equivalent between FBP and IR, whereas CNR with IR was significantly higher than that with FBP. %GGO significantly increased at low-dose levels with FBP; however, IR suppressed the elevation.

Conclusion: The confidence ratings of diagnoses of simulated lesions other than lung cancer on low-dose CT screening were not degraded with hybrid IR compared with FBP.

Advances In Knowledge: Hybrid IR did not degrade the confidence ratings of diagnoses on visual assessment and differential diagnoses based on CT value of mediastinal masses, and it showed the advantage of higher GGO conspicuity at low-dose level. Radiologists can analyse images of hybrid IR alone on low-dose CT screening for lung cancer.

Citing Articles

Low-dose CT in COVID-19 outbreak: radiation safety, image wisely, and image gently pledge.

Tofighi S, Najafi S, Johnston S, Gholamrezanezhad A Emerg Radiol. 2020; 27(6):601-605.

PMID: 32390122 PMC: 7211266. DOI: 10.1007/s10140-020-01784-3.

References
1.
Gervaise A, Osemont B, Lecocq S, Noel A, Micard E, Felblinger J . CT image quality improvement using Adaptive Iterative Dose Reduction with wide-volume acquisition on 320-detector CT. Eur Radiol. 2011; 22(2):295-301. DOI: 10.1007/s00330-011-2271-7. View

2.
W Wille M, Thomsen L, Petersen J, De Bruijne M, Dirksen A, Pedersen J . Visual assessment of early emphysema and interstitial abnormalities on CT is useful in lung cancer risk analysis. Eur Radiol. 2015; 26(2):487-94. DOI: 10.1007/s00330-015-3826-9. View

3.
Richard S, Husarik D, Yadava G, Murphy S, Samei E . Towards task-based assessment of CT performance: system and object MTF across different reconstruction algorithms. Med Phys. 2012; 39(7):4115-22. DOI: 10.1118/1.4725171. View

4.
Nishio M, Matsumoto S, Ohno Y, Sugihara N, Inokawa H, Yoshikawa T . Emphysema quantification by low-dose CT: potential impact of adaptive iterative dose reduction using 3D processing. AJR Am J Roentgenol. 2012; 199(3):595-601. DOI: 10.2214/AJR.11.8174. View

5.
Ohno Y, Takenaka D, Kanda T, Yoshikawa T, Matsumoto S, Sugihara N . Adaptive iterative dose reduction using 3D processing for reduced- and low-dose pulmonary CT: comparison with standard-dose CT for image noise reduction and radiological findings. AJR Am J Roentgenol. 2012; 199(4):W477-85. DOI: 10.2214/AJR.11.8275. View