» Articles » PMID: 25976600

The Precision of Pulmonary Artery Catheter Bolus Thermodilution Cardiac Output Measurements Varies With the Clinical Situation

Overview
Date 2015 May 16
PMID 25976600
Citations 9
Authors
Affiliations
Soon will be listed here.
Abstract

Objective: To investigate the effects of ventilatory mode, injectate temperature, and clinical situation on the precision of cardiac output measurements.

Design: Randomized, prospective observational study.

Setting: Single university hospital.

Participants: Forty patients undergoing planned cardiac surgery, receiving a pulmonary artery catheter according to institutional routine.

Interventions: Cardiac output was measured at 4 predefined time points during the perioperative patient course, twice during controlled and twice during spontaneous ventilation, using 2 blocks of 8 measurement replications with cold and tepid injectate in random order.

Measurements And Main Results: The data were analyzed using a hierarchical linear mixed model. Clinical precision was determined as half the width of the 95% confidence interval for the underlying true value. The single-measurement precision measured in 2 different clinical situations for each temperature/ventilation combination was 8% to 10%, 11% to 13%, 13% to 15%, and 23% to 24% in controlled ventilation with cold injectate, controlled ventilation with tepid injectate, spontaneous breathing with cold injectate, and spontaneous breathing with tepid injectate, respectively. Tables are provided for the number of replications needed to achieve a certain precision and for how to identify significant changes in cardiac output.

Conclusions: Clinical precision of cardiac output measurements is reduced significantly during spontaneous relative to controlled ventilation. The differences in precision between repeated measurement series within the temperature/ventilation combinations indicate influence of other situation-specific factors not related to ventilatory mode. Compared with tepid injectate in patients breathing spontaneously, the precision is 3-fold better with cold injectate and controlled ventilation.

Citing Articles

Prediction of fluid responsiveness in spontaneously breathing patients with hemodynamic stability: a prospective repeated-measures study.

Kim Y, Lee J Sci Rep. 2024; 14(1):14451.

PMID: 38914634 PMC: 11196262. DOI: 10.1038/s41598-024-65554-8.


Accuracy of a noninvasive estimated continuous cardiac output measurement under different respiratory conditions: a prospective observational study.

Takakura M, Fujii T, Taniguchi T, Suzuki S, Nishiwaki K J Anesth. 2023; 37(3):394-400.

PMID: 36905408 DOI: 10.1007/s00540-023-03176-6.


Severe loss of right ventricular longitudinal contraction occurs after cardiopulmonary bypass in patients with preserved right ventricular output.

Gronlykke L, Korshin A, Holmgaard F, Kjoller S, Gustafsson F, Nilsson J Int J Cardiovasc Imaging. 2019; 35(9):1661-1670.

PMID: 31049752 DOI: 10.1007/s10554-019-01616-7.


Post cardiac surgery stunning reduces stroke work, but leaves cardiac power output unchanged in patients with normal ejection fraction.

Tannvik T, Rimehaug A, Skjaervold N, Kirkeby-Garstad I Physiol Rep. 2018; 6(13):e13781.

PMID: 29998610 PMC: 6041697. DOI: 10.14814/phy2.13781.


Performance of a capnodynamic method estimating effective pulmonary blood flow during transient and sustained hypercapnia.

Sigmundsson T, Ohman T, Hallback M, Redondo E, Suarez Sipmann F, Wallin M J Clin Monit Comput. 2017; 32(2):311-319.

PMID: 28497180 PMC: 5838142. DOI: 10.1007/s10877-017-0021-3.