» Articles » PMID: 25890013

A Comparative Assessment of Prostate Positioning Guided by Three-dimensional Ultrasound and Cone Beam CT

Overview
Journal Radiat Oncol
Publisher Biomed Central
Specialties Oncology
Radiology
Date 2015 Apr 19
PMID 25890013
Citations 19
Authors
Affiliations
Soon will be listed here.
Abstract

Background: The accuracy of the Elekta Clarity™ three-dimensional ultrasound system (3DUS) was assessed for prostate positioning and compared to seed- and bone-based positioning in kilo-voltage cone-beam computed tomography (CBCT) during a definitive radiotherapy.

Methods: The prostate positioning of 6 patients, with fiducial markers implanted into the prostate, was controlled by 3DUS and CBCT. In total, 78 ultrasound scans were performed trans-abdominally and compared to bone-matches and seed-matches in CBCT scans. Setup errors detected by the different modalities were compared. Systematic and random errors were analysed, and optimal setup margins were calculated.

Results: The discrepancy between 3DUS and seed-match in CBCT was -0.2 ± 2.7 mm laterally, -1.9 ± 2.3 mm longitudinally and 0.0 ± 3.0 mm vertically and significant only in longitudinal direction. Using seed-match as reference, systematic errors of 3DUS were 1.3 mm laterally, 0.8 mm longitudinally and 1.4 mm vertically, and random errors were 2.5 mm laterally, 2.3 mm longitudinally, and 2.7 mm vertically. No significant difference could be detected for 3DUS in comparison to bone-match in CBCT.

Conclusions: 3DUS is feasible for image guidance for patients with prostate cancer and appears comparable to CBCT based image guidance in the retrospective study. While 3DUS offers some distinct advantages such as no need of invasive fiducial implantation and avoidance of extra radiation, its disadvantages include the operator dependence of the technique and dependence on sufficient bladder filling. Further study of 3DUS for image guidance in a large patient cohort is warranted.

Citing Articles

Deep learning based ultra-low dose fan-beam computed tomography image enhancement algorithm: Feasibility study in image quality for radiotherapy.

Jiang H, Qin S, Jia L, Wei Z, Xiong W, Xu W J Appl Clin Med Phys. 2024; 25(12):e14560.

PMID: 39540681 PMC: 11633815. DOI: 10.1002/acm2.14560.


Physician reported toxicities and patient reported quality of life of transperineal ultrasound-guided radiotherapy of prostate cancer.

Ma J, Rogowski P, Trapp C, Manapov F, Xu B, Buchner A Clin Transl Radiat Oncol. 2024; 49:100868.

PMID: 39381629 PMC: 11459703. DOI: 10.1016/j.ctro.2024.100868.


Optical Surface Management System and BladderScan for Patient Setup During Radiotherapy of Postoperative Prostate Cancer.

Chen H, Liu Y, Qin S, Gan G Biomed Res Int. 2024; 2024:3573796.

PMID: 39263420 PMC: 11390218. DOI: 10.1155/2024/3573796.


Deep learning-based low-dose CT for adaptive radiotherapy of abdominal and pelvic tumors.

Gong W, Yao Y, Ni J, Jiang H, Jia L, Xiong W Front Oncol. 2022; 12:968537.

PMID: 36059630 PMC: 9436420. DOI: 10.3389/fonc.2022.968537.


Dosimetric Impact of Inter-Fraction Anatomical Changes in Carbon Ion Boost Treatment for High-Risk Prostate Cancer (AIRC IG 14300).

Russo S, Ricotti R, Molinelli S, Patti F, Barcellini A, Mastella E Front Oncol. 2021; 11:740661.

PMID: 34650922 PMC: 8506150. DOI: 10.3389/fonc.2021.740661.


References
1.
Ali A, Dirkx M, Cools R, Heijmen B . Accurate IMRT fluence verification for prostate cancer patients using 'in-vivo' measured EPID images and in-room acquired kilovoltage cone-beam CT scans. Radiat Oncol. 2013; 8:211. PMC: 3851603. DOI: 10.1186/1748-717X-8-211. View

2.
Wu Q, Li T, Yuan L, Yin F, Lee W . Single institution's dosimetry and IGRT analysis of prostate SBRT. Radiat Oncol. 2013; 8:215. PMC: 3852330. DOI: 10.1186/1748-717X-8-215. View

3.
van Herk M, Remeijer P, Rasch C, Lebesque J . The probability of correct target dosage: dose-population histograms for deriving treatment margins in radiotherapy. Int J Radiat Oncol Biol Phys. 2000; 47(4):1121-35. DOI: 10.1016/s0360-3016(00)00518-6. View

4.
Pollack A, Zagars G, Starkschall G, Antolak J, Lee J, Huang E . Prostate cancer radiation dose response: results of the M. D. Anderson phase III randomized trial. Int J Radiat Oncol Biol Phys. 2002; 53(5):1097-105. DOI: 10.1016/s0360-3016(02)02829-8. View

5.
Chandra A, Dong L, Huang E, Kuban D, ONeill L, Rosen I . Experience of ultrasound-based daily prostate localization. Int J Radiat Oncol Biol Phys. 2003; 56(2):436-47. DOI: 10.1016/s0360-3016(02)04612-6. View