Can Larger-bodied Cemented Femoral Components Reduce Periprosthetic Fractures? A Biomechanical Study
Overview
General Surgery
Orthopedics
Affiliations
Introduction: The risk for late periprosthetic femoral fractures is higher in patients treated for a neck of femur fracture compared to osteoarthritis. It has been hypothesised that osteopaenia and consequent decreased stiffness of the proximal femur are responsible for this. We investigated whether a femoral component with a bigger body would increase the torque to failure in a biaxially loaded composite Sawbone model.
Materials And Methods: A biomechanical bone analogue was used. Two different body sizes (Exeter 44-1 versus 44-4) of a polished tapered cemented femoral stem were implanted by an experienced surgeon in seven bone analogues each and internally rotated at 40°/s until failure. Torque to fracture and fracture energy were measured using a biaxial materials testing device (Instron 8874, MI, USA). The data were non-parametric and therefore tested with the Mann-Whitney U test.
Results: The median torque to fracture was 156.7 Nm (IQR 19.7) for the 44-1 stem and 237.1 Nm (IQR 52.9) for the 44-4 stem (p = 0.001). The median fracture energy was 8.5 J (IQR 7.3) for the 44-1 stem and 19.5 J (IQR 8.8) for the 44-4 stem (p = 0.014).
Conclusion: The use of large body polished tapered cemented stems for neck of femur fractures increases the torque to failure in a biomechanical model and therefore is likely to reduce late periprosthetic fracture risk in this vulnerable cohort.
Lamb J, Coltart O, Adekanmbi I, Pandit H, Stewart T Proc Inst Mech Eng H. 2022; 236(7):973-978.
PMID: 35603754 PMC: 9289970. DOI: 10.1177/09544119221092842.
Takegami Y, Seki T, Osawa Y, Imagama S Bone Joint Res. 2022; 11(5):270-277.
PMID: 35514237 PMC: 9130673. DOI: 10.1302/2046-3758.115.BJR-2021-0375.R2.
Irie T, Takahashi D, Asano T, Shimizu T, Arai R, Terkawi A BMC Musculoskelet Disord. 2019; 20(1):355.
PMID: 31370807 PMC: 6670119. DOI: 10.1186/s12891-019-2716-8.
Cementing of the hip arthroplasty stem increases load-to-failure force: a cadaveric study.
Klasan A, Baumlein M, Bliemel C, Putnis S, Neri T, Schofer M Acta Orthop. 2019; 90(5):445-449.
PMID: 31282247 PMC: 6746255. DOI: 10.1080/17453674.2019.1634331.
Stevens J, Clement N, Nasserallah M, Millar M, Joseph S Eur J Orthop Surg Traumatol. 2018; 28(4):659-665.
PMID: 29423864 DOI: 10.1007/s00590-018-2141-y.