Inter-operator and Inter-device Agreement and Reliability of the SEM Scanner

Overview
Specialties Nursing
Physiology
Date 2015 Feb 16
PMID 25682271
Citations 11
Authors
Affiliations
Soon will be listed here.
Abstract

Objective: The SEM Scanner is a medical device designed for use by healthcare providers as part of pressure ulcer prevention programs. The objective of this study was to evaluate the inter-rater and inter-device agreement and reliability of the SEM Scanner.

Methods: Thirty-one (31) volunteers free of pressure ulcers or broken skin at the sternum, sacrum, and heels were assessed with the SEM Scanner. Each of three operators utilized each of three devices to collect readings from four anatomical sites (sternum, sacrum, left and right heels) on each subject for a total of 108 readings per subject collected over approximately 30 min. For each combination of operator-device-anatomical site, three SEM readings were collected. Inter-operator and inter-device agreement and reliability were estimated.

Results: Over the course of this study, more than 3000 SEM Scanner readings were collected. Agreement between operators was good with mean differences ranging from -0.01 to 0.11. Inter-operator and inter-device reliability exceeded 0.80 at all anatomical sites assessed.

Conclusion: The results of this study demonstrate the high reliability and good agreement of the SEM Scanner across different operators and different devices. Given the limitations of current methods to prevent and detect pressure ulcers, the SEM Scanner shows promise as an objective, reliable tool for assessing the presence or absence of pressure-induced tissue damage such as pressure ulcers.

Citing Articles

A systematic review on the impact of sub-epidermal moisture assessments on pressure ulcer/injury care delivery pathways.

Avsar P, Patton D, Cuddigan J, Moore Z Int Wound J. 2024; 21(6):e14928.

PMID: 38832363 PMC: 11148479. DOI: 10.1111/iwj.14928.


Inter-device agreement of sacral subepidermal oedema measurement in healthy adults during prolonged 60° head of bed elevation.

Latimer S, Bone M, Walker R, Thalib L, Gillespie B Nurs Open. 2024; 11(2):e2103.

PMID: 38391104 PMC: 10830921. DOI: 10.1002/nop2.2103.


Sub epidermal moisture measurement and targeted SSKIN bundle interventions, a winning combination for the treatment of early pressure ulcer development.

Byrne S, Patton D, Avsar P, Strapp H, Budri A, OConnor T Int Wound J. 2022; 20(6):1987-1999.

PMID: 36575149 PMC: 10333035. DOI: 10.1111/iwj.14061.


Imaging in Chronic Wound Diagnostics.

Li S, Mohamedi A, Senkowsky J, Nair A, Tang L Adv Wound Care (New Rochelle). 2020; 9(5):245-263.

PMID: 32226649 PMC: 7099416. DOI: 10.1089/wound.2019.0967.


A blinded clinical study using a subepidermal moisture biocapacitance measurement device for early detection of pressure injuries.

Okonkwo H, Bryant R, Milne J, Molyneaux D, Sanders J, Cunningham G Wound Repair Regen. 2020; 28(3):364-374.

PMID: 31965682 PMC: 7217158. DOI: 10.1111/wrr.12790.