» Articles » PMID: 25659140

A Comparison Between Conductive and Infrared Devices for Measuring Mean Skin Temperature at Rest, During Exercise in the Heat, and Recovery

Overview
Journal PLoS One
Date 2015 Feb 7
PMID 25659140
Citations 17
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: Skin temperature assessment has historically been undertaken with conductive devices affixed to the skin. With the development of technology, infrared devices are increasingly utilised in the measurement of skin temperature. Therefore, our purpose was to evaluate the agreement between four skin temperature devices at rest, during exercise in the heat, and recovery.

Methods: Mean skin temperature ([Formula: see text]) was assessed in thirty healthy males during 30 min rest (24.0 ± 1.2°C, 56 ± 8%), 30 min cycle in the heat (38.0 ± 0.5°C, 41 ± 2%), and 45 min recovery (24.0 ± 1.3°C, 56 ± 9%). [Formula: see text] was assessed at four sites using two conductive devices (thermistors, iButtons) and two infrared devices (infrared thermometer, infrared camera).

Results: Bland-Altman plots demonstrated mean bias ± limits of agreement between the thermistors and iButtons as follows (rest, exercise, recovery): -0.01 ± 0.04, 0.26 ± 0.85, -0.37 ± 0.98°C; thermistors and infrared thermometer: 0.34 ± 0.44, -0.44 ± 1.23, -1.04 ± 1.75°C; thermistors and infrared camera (rest, recovery): 0.83 ± 0.77, 1.88 ± 1.87°C. Pairwise comparisons of [Formula: see text] found significant differences (p < 0.05) between thermistors and both infrared devices during resting conditions, and significant differences between the thermistors and all other devices tested during exercise in the heat and recovery.

Conclusions: These results indicate poor agreement between conductive and infrared devices at rest, during exercise in the heat, and subsequent recovery. Infrared devices may not be suitable for monitoring [Formula: see text] in the presence of, or following, metabolic and environmental induced heat stress.

Citing Articles

Non-Contact Infrared Thermometers and Thermal Scanners for Human Body Temperature Monitoring: A Systematic Review.

Zhao Y, Bergmann J Sensors (Basel). 2023; 23(17).

PMID: 37687902 PMC: 10490756. DOI: 10.3390/s23177439.


Swimming performance, physiology, and post-activation performance enhancement following dryland transition phase warmup: A systematic review.

McKenzie M, McKean M, Doyle D, Hogarth L, Burkett B PLoS One. 2022; 17(8):e0273248.

PMID: 35980952 PMC: 9387820. DOI: 10.1371/journal.pone.0273248.


A Thermal Skin Model for Comparing Contact Skin Temperature Sensors and Assessing Measurement Errors.

MacRae B, Spengler C, Psikuta A, Rossi R, Annaheim S Sensors (Basel). 2021; 21(14).

PMID: 34300649 PMC: 8309895. DOI: 10.3390/s21144906.


Flexible Hybrid Electrodes for Continuous Measurement of the Local Temperature in Long-Term Wounds.

Rodes-Carbonell A, Torregrosa-Valls J, Guill Ibanez A, Tormos Ferrando A, Juan Blanco M, Ferriols A Sensors (Basel). 2021; 21(8).

PMID: 33924527 PMC: 8070585. DOI: 10.3390/s21082741.


Recent Advances on Early Detection of Heat Strain in Dairy Cows Using Animal-Based Indicators: A Review.

Shu H, Wang W, Guo L, Bindelle J Animals (Basel). 2021; 11(4).

PMID: 33915761 PMC: 8066310. DOI: 10.3390/ani11040980.


References
1.
Cheung S, McLellan T, Tenaglia S . The thermophysiology of uncompensable heat stress. Physiological manipulations and individual characteristics. Sports Med. 2000; 29(5):329-59. DOI: 10.2165/00007256-200029050-00004. View

2.
Wilmore D, MASON Jr A, Johnson D, PRUITT Jr B . Effect of ambient temperature on heat production and heat loss in burn patients. J Appl Physiol. 1975; 38(4):593-7. DOI: 10.1152/jappl.1975.38.4.593. View

3.
Niu H, Lui P, Hu J, Ting C, Yin Y, Lo Y . Thermal symmetry of skin temperature: normative data of normal subjects in Taiwan. Zhonghua Yi Xue Za Zhi (Taipei). 2001; 64(8):459-68. View

4.
Wasner G, Schattschneider J, Baron R . Skin temperature side differences--a diagnostic tool for CRPS?. Pain. 2002; 98(1-2):19-26. DOI: 10.1016/s0304-3959(01)00470-5. View

5.
van den Heuvel C, Ferguson S, Dawson D, Gilbert S . Comparison of digital infrared thermal imaging (DITI) with contact thermometry: pilot data from a sleep research laboratory. Physiol Meas. 2003; 24(3):717-25. DOI: 10.1088/0967-3334/24/3/308. View