Abstract

Background: Fiberoptical assisted intubation via a placed laryngeal mask airway (LMA) has been described as save and easy procedure to manage a difficult airway. The laryngeal tube (LT) is a promising alternative to the LMA as supraglottic airway device. Fiberoptical assisted intubation via LT is possible, however considered more difficult. The aim of this study was to compare the fiberoptical assisted intubation via LT and LMA.

Materials And Methods: A total of 22 anesthesiologists with different levels of experience participated in the study performed on an adult airway model. Primarily the supraglottic device was placed and correct position was confirmed by successful ventilation. A 5 mm internal diameter tracheal tube was loaded onto a flexible 3.6 mm fiberscope and the so prepared device was inserted into the proximal lumen of the LMA or the LT. The glottis was passed under visual control and the tube advanced into the trachea. After removal of the fiberscope, ventilation was examined clinically by inspection. Success rates, procedure time and observed complications of LMA versus LT were compared (U-test; P < 0.05).

Results: Placement of the endotracheal tube was successful in all attempts using both the LMA and LT. There was no difference in the time needed for the placement procedure (33 [26-38] s LMA; 35 [32-38] s LT). Only minor technical complications were observed in both groups.

Conclusion: A fiberoptical assisted intubation via LT can be considered as a relevant alternative in advanced airway management.

Citing Articles

[Intubating laryngeal tube suction disposable: Initial clinical experiences with a novel device for endotracheal intubation].

Bergold M, Kahle S, Schultzik T, Bucheler M, Byhahn C Anaesthesist. 2015; 65(1):30-35.

PMID: 26481391 DOI: 10.1007/s00101-015-0100-0.

References
1.
Charters P, OSullivan E . The 'dedicated airway': a review of the concept and an update of current practice. Anaesthesia. 1999; 54(8):778-86. DOI: 10.1046/j.1365-2044.1999.00888.x. View

2.
Asai T, Murao K, Shingu K . Efficacy of the laryngeal tube during intermittent positive-pressure ventilation. Anaesthesia. 2000; 55(11):1099-102. DOI: 10.1046/j.1365-2044.2000.01710.x. View

3.
Atherton D, OSullivan E, Lowe D, Charters P . A ventilation-exchange bougie for fibreoptic intubations with the laryngeal mask airway. Anaesthesia. 1996; 51(12):1123-6. DOI: 10.1111/j.1365-2044.1996.tb15047.x. View

4.
Richebe P, Semjen F, Cros A, Maurette P . Clinical assessment of the laryngeal tube in pediatric anesthesia. Paediatr Anaesth. 2005; 15(5):391-6. DOI: 10.1111/j.1460-9592.2005.01469.x. View

5.
Cook T, Hardy R, McKinstry C, Twigg S . Use of the laryngeal tube as a dedicated airway during tracheal intubation. Br J Anaesth. 2003; 90(3):397-9; author reply 397-9. DOI: 10.1093/bja/aeg538. View