Resection of Calcified Aortic Heart Leaflets in Vitro by Q-switched 2 µm Microsecond Laser Radiation
Overview
General Surgery
Affiliations
Background: Transcatheter aortic valve implantation (TAVI) can result in paravalvular leakage and stent deformation in the presence of severe calcification. This study was undertaken to determine the efficacy of laser-assisted resection of calcific aortic valve leaflets as a method to minimize the effects of calcium on perivalvular leakage during TAVI.
Methods: A Q-switched Tm:YAG laser emitting at a wavelength of 2.01 μm was used to evaluate the cutting efficiency on highly calcified human aortic leaflets in vitro (N = 10). A pulse energy of 4.3 mJ, a pulse duration of 0.8-1 μs, and a repetition rate of 1 kHz were used. The radiation was transmitted via a 200 µm core diameter quartz fiber. Resection was performed in a fiber-tissue contact mode on water-covered samples in a dish. The remnant particles were analyzed with respect to quantity and size by light microscopy.
Results: A resection rate of 40.4 ± 22.2 mg/min on highly calcified aortic leaflets was achieved. This corresponds to a cutting speed of approximately 1 cm/min; a laser dissection time of 3 min per leaflet is expected. The majority of the remnant particles (85.4%) were <6 μm in diameter, with only 0.1% exceeding 300 μm.
Conclusions: The Q-switched Tm:YAG laser system showed promising results in cutting calcified aortic valves, by transmitting sufficient energy through a small flexible fiber. Catheter-based removal of aortic valve calcification may help to improve TAVI technology.
Ventimiglia E, Robesti D, Bevilacqua L, Tondelli E, Oliva I, Orecchia L World J Urol. 2023; 41(11):3301-3308.
PMID: 37682286 DOI: 10.1007/s00345-023-04580-z.
Effect of laser pulse shaping on photoacoustic dosimetry in retinal models.
Brown R, Dufour S, Deladurantaye P, Bouch N, Gallant P, Methot S Biomed Opt Express. 2020; 11(11):6590-6604.
PMID: 33282510 PMC: 7687941. DOI: 10.1364/BOE.403703.
Transcatheter aortic valve resection: new mechanical devices.
Dai H, Lutter G, Frank D, Freitag-Wolf S, Topal A, Haneya A J Thorac Dis. 2020; 12(11):6586-6597.
PMID: 33282360 PMC: 7711357. DOI: 10.21037/jtd-20-2036.