Comparative Analysis of Corneal Measurements Obtained from a Scheimpflug Camera and an Integrated Placido-optical Coherence Tomography Device in Normal and Keratoconic Eyes

Overview
Journal Acta Ophthalmol
Specialty Ophthalmology
Date 2014 Dec 16
PMID 25495530
Citations 16
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: To assess the agreement between a Scheimpflug camera (Pentacam) and a combined Placido-optical coherence tomography device (Visante OMNI) in measuring corneal curvature, thickness and elevation values in normal and keratoconic eyes.

Methods: Corneal measurements of 110 normal eyes (one eye per subject) and 70 keratoconic eyes were obtained from both devices and compared. Agreement was determined using the Bland-Altman analysis 95% limits of agreement (LoA).

Results: The Pentacam measured significantly greater keratometry readings in the flattest (K1) and steepest meridians (K2) in normal and keratoconic eyes. The 95% LoA in normal eyes were -0.32 to 0.59 dioptres (D) (K1) and -0.41 to 0.74 D (K2). In keratoconic eyes, the 95% LoA were -1.35 to 1.92 D (K1) and -1.38 to 1.99 D (K2). The Pentacam recorded significantly higher central corneal thickness (CCT) values in both groups of eyes. The 95% LoA were -4.31 to 39.89 microns (μ) and -12.92 to 41.35 μ in normal and keratoconic eyes, respectively. Pentacam anterior and posterior corneal elevations were significantly greater in both groups of eyes. The devices demonstrated excellent repeatability and reproducibility for corneal curvature and thickness but not elevation measurements.

Conclusions: The Pentacam measured significantly greater corneal curvature, thickness and elevation values compared to the Visante OMNI in normal and keratoconic eyes. The devices agree moderately for anterior corneal elevations in normal eyes and do not appear to be interchangeable for corneal measurements in clinical practice.

Citing Articles

Comparison of a Scheimpflug Camera and Optical Coherence Tomography in Evaluating Keratoconic Eyes Post Keratoplasty.

Gadamer A, Miklaszewski P, Janiszewska-Bil D, Lyssek-Boron A, Dobrowolski D, Wylegala E J Clin Med. 2025; 14(1.

PMID: 39797319 PMC: 11722208. DOI: 10.3390/jcm14010238.


Reduced Retinal Blood Vessel Densities Measured by Optical Coherence Tomography Angiography in Keratoconus Patients Are Negatively Correlated with Keratoconus Severity.

Leclaire M, Storp J, Lahme L, Esser E, Eter N, Alnawaiseh M Diagnostics (Basel). 2024; 14(7).

PMID: 38611620 PMC: 11011292. DOI: 10.3390/diagnostics14070707.


Posterior corneal elevation changes during 12 month of overnight orthokeratology.

Wang Z, Wang Z, Meng Y, Wang P, Yibulayin S, Jiang B Heliyon. 2023; 9(4):e14887.

PMID: 37025848 PMC: 10070639. DOI: 10.1016/j.heliyon.2023.e14887.


Comparative Analysis of Corneal Parameters Performed with GalileiG6 and OCT Casia 2.

Mazur R, Wylegala A, Wylegala E, Dobrowolski D Diagnostics (Basel). 2023; 13(2).

PMID: 36673077 PMC: 9857403. DOI: 10.3390/diagnostics13020267.


Comparison between the CASIA SS-1000 and Pentacam in measuring corneal curvatures and corneal thickness maps.

Feldman R, Kim G, Chuang A, Shiraishi A, Okamoto K, Tsukamoto M BMC Ophthalmol. 2023; 23(1):10.

PMID: 36604657 PMC: 9814456. DOI: 10.1186/s12886-023-02768-w.