Persistent Blood Stream Infection in Patients Supported with a Continuous-flow Left Ventricular Assist Device is Associated with an Increased Risk of Cerebrovascular Accidents

Abstract

Background: Common adverse events in patients supported with Continuous-flow left ventricular assist devices (CF-LVAD) include infections and cerebrovascular accidents (CVA). Some studies have suggested a possible association between blood stream infection (BSI) and CVA.

Methods And Results: Medical records of patients who received Heartmate II (HMII) CF-LVADs in 2008-2012 at a single center were reviewed. CVA was categorized as either hemorrhagic (HCVA) or ischemic (ICVA). BSI was divided into persistent (pBSI) and nonpersistent (non-pBSI). pBSI was defined as BSI with the same organism on repeated blood culture >72 hours from initial blood culture despite antibiotics. Univariate and multivariate analyses were performed to determine predictors. A total of 149 patients had HMII implanted; 76% were male, and the overall mean age was 55.4 ± 13 years. There were a total of 19 (13%) patients who had CVA (7 HCVA and 12 ICVA) at a median of 295 days (range 5-1,096 days) after implantation. There were a total of 28 (19%) patients with pBSI and 17 (11%) patients with non-pBSI. Patients with pBSI had a trend toward greater BMI (31 kg/m(2) vs 27 kg/m(2); P = .09), and longer duration of support (1,019 d vs 371 d; P < .001) compared with those with non-pBSI. Persistent BSI was associated with an increased risk of mortality and with all-cause CVA on multivariate analysis (odds ratio [OR] 5.97; P = .003) as well as persistent Pseudomonas aeruginosa infection (OR 4.54; P = .048).

Conclusions: Persistent BSI is not uncommon in patients supported by CF-LVAD and is highly associated with all-cause CVA and increased all-cause mortality.

Citing Articles

F-FDG PET/CT in left ventricular assist device infections: In-depth characterization and clinical implications.

Devesa A, Rashed E, Moss N, Robson P, Pyzik R, Roldan J J Heart Lung Transplant. 2023; 43(4):529-538.

PMID: 37951322 PMC: 10969750. DOI: 10.1016/j.healun.2023.11.002.


Hypoxic-Ischemic Brain Injury in ECMO: Pathophysiology, Neuromonitoring, and Therapeutic Opportunities.

Khanduja S, Kim J, Kang J, Feng C, Vogelsong M, Geocadin R Cells. 2023; 12(11).

PMID: 37296666 PMC: 10252448. DOI: 10.3390/cells12111546.


Added value of semi-quantitative analysis of [18F]FDG PET/CT for the diagnosis of device-related infections in patients with a left ventricular assist device.

Ten Hove D, Wahadat A, Slart R, Wouthuyzen-Bakker M, Mecozzi G, Damman K Eur Heart J Cardiovasc Imaging. 2022; 24(6):819-828.

PMID: 36573930 PMC: 10229264. DOI: 10.1093/ehjci/jeac260.


Neurologic Complications in Patients With Left Ventricular Assist Devices.

Shoskes A, Fan T, Starling R, Cho S Can J Cardiol. 2022; 39(2):210-221.

PMID: 36400374 PMC: 9905352. DOI: 10.1016/j.cjca.2022.11.004.


Endovascular therapy for intracranial infectious aneurysms associated with a left ventricular assist device: illustrative case.

Okuda T, Nishimura A, Arimura K, Iwaki K, Fujino T, Ushijima T J Neurosurg Case Lessons. 2022; 3(11).

PMID: 36209407 PMC: 9379626. DOI: 10.3171/CASE21559.