Plate Selection for Fixation of Extra-articular Distal Humerus Fractures: a Biomechanical Comparison of Three Different Implants

Overview
Journal Injury
Publisher Elsevier
Specialty Emergency Medicine
Date 2014 Sep 25
PMID 25249244
Citations 21
Authors
Affiliations
Soon will be listed here.
Abstract

Operative fixation of extra-articular distal humerus using a single posterolateral column plate has been described but the biomechanical properties or limits of this technique is undefined. The purpose of this study was to evaluate the mechanical properties of distal humerus fracture fixation using three standard fixation constructs. Two equal groups were created from forty sawbones humeri. Osteotomies were created at 80mm or 50mm from the tip of the trochlea. In the proximal osteotomy group, sawbones were fixed with an 8-hole 3.5mm LCP or with a 6-hole posterolateral plate. In the distal group, sawbones were fixed with 9-hole medial and lateral 3.5mm distal humerus plates and ten sawbones were fixed with a 6-hole posterolateral plate. Biomechanical testing was performed using a servohydraulic testing machine. Testing in extension as well as internal and external rotation was performed. Destructive testing was also performed with failure being defined as hardware pullout, sawbone failure or cortical contact at the osteotomy. In the proximal osteotomy group, the average bending stiffness and torsional stiffness was significantly greater with the posterolateral plate than with the 3.5mm LCP. In the distal osteotomy group, the average bending stiffness and torsional stiffness was significantly greater with the posterolateral plate than the 3.5mm LCP. In extension testing, the yield strength was significantly greater with the posterolateral plate in the proximal osteotomy specimens, and the dual plating construct in the distal osteotomy specimens. The yield strength of specimens in axial torsion was significantly greater with the posterolateral plate in the proximal osteotomy specimens, and the dual plating construct in the distal osteotomy specimens. Limited biomechanical data to support the use of a pre-contoured posterolateral distal humerus LCP for fixation of extra-articular distal humerus exists. We have found that this implant provided significantly greater bending stiffness, torsional stiffness, and yield strength than a single 3.5mm LCP plate for osteotomies created 80mm from the trochlea. At the more distal osteotomy, dual plating was biomechanically superior. Our results suggest that single posterolateral column fixation of extra-articular humerus fractures is appropriate for more proximal fractures but that dual plate fixation is superior for more distal fractures.

Citing Articles

The clinical outcomes of anterior minimally invasive inverted PHILOS plate fixation for distal humeral shaft fractures.

Fu G, Wang S, Wu W, Lin F, Li R J Orthop Surg Res. 2025; 20(1):177.

PMID: 39972364 PMC: 11837353. DOI: 10.1186/s13018-025-05560-2.


Biomechanical Design Optimization of Distal Humerus Fracture Plates: A Review.

Zdero R, Brzozowski P, Schemitsch E Biomed Res Int. 2024; 2024:6015794.

PMID: 38966093 PMC: 11223906. DOI: 10.1155/2024/6015794.


Treatment of distal third humeral shaft fractures with posterior minimally invasive plate osteosynthesis (MIPO) with segmental isolation of the radial nerve: minimum one-year follow-up.

Contreras J, Soto D, Valencia M, Lopez M, Diaz A, Delgado S JSES Rev Rep Tech. 2024; 4(1):53-60.

PMID: 38323209 PMC: 10840569. DOI: 10.1016/j.xrrt.2023.08.006.


Treatment of Distal Third Humeral Shaft Fracture with Intramedullary Nail Combined with Anterior Minimally Invasive Plate Osteosynthesis.

Fu G, Zhang Y, Ke S, Zhu D, Wu J, Su D Orthop Surg. 2023; 15(12):3101-3107.

PMID: 37817420 PMC: 10693993. DOI: 10.1111/os.13893.


Biomechanical performance evaluation of a modified proximal humerus locking plate for distal humerus shaft fracture using finite element analysis.

Lee J, Kim K, Yoon Y Sci Rep. 2023; 13(1):16250.

PMID: 37758839 PMC: 10533898. DOI: 10.1038/s41598-023-43183-x.