We demonstrate the feasibility of microscale molecular imaging using hyperpolarized proton and carbon-13 MRI contrast media and low-field (47.5 mT) preclinical scale (38 mm i.d.) 2D magnetic resonance imaging (MRI). Hyperpolarized proton images with 94 × 94 μm(2) spatial resolution and hyperpolarized carbon-13 images with 250 × 250 μm(2) in-plane spatial resolution were recorded in 4-8 s (largely limited by the electronics response), surpassing the in-plane spatial resolution (i.e., pixel size) achievable with micro-positron emission tomography (PET). These hyperpolarized proton and (13)C images were recorded using large imaging matrices of up to 256 × 256 pixels and relatively large fields of view of up to 6.4 × 6.4 cm(2). (13)C images were recorded using hyperpolarized 1-(13)C-succinate-d2 (30 mM in water, %P(13C) = 25.8 ± 5.1% (when produced) and %P(13C) = 14.2 ± 0.7% (when imaged), T1 = 74 ± 3 s), and proton images were recorded using (1)H hyperpolarized pyridine (100 mM in methanol-d4, %P(H) = 0.1 ± 0.02% (when imaged), T1 = 11 ± 0.1 s). Both contrast agents were hyperpolarized using parahydrogen (>90% para-fraction) in an automated 5.75 mT parahydrogen induced polarization (PHIP) hyperpolarizer. A magnetized path was demonstrated for successful transportation of a (13)C hyperpolarized contrast agent (1-(13)C-succinate-d2, sensitive to fast depolarization when at the Earth's magnetic field) from the PHIP polarizer to the 47.5 mT low-field MRI. While future polarizing and low-field MRI hardware and imaging sequence developments can further improve the low-field detection sensitivity, the current results demonstrate that microscale molecular imaging in vivo is already feasible at low (<50 mT) fields and potentially at low (~1 mM) metabolite concentrations.

Citing Articles

Low-Cost Purpose-Built Ultra-Low-Field NMR Spectrometer.

Chowdhury M, Ahmed F, Oladun C, Adelabu I, Abdurraheem A, Nantogma S Anal Chem. 2024; 96(42):16724-16734.

PMID: 39378166 PMC: 11506762. DOI: 10.1021/acs.analchem.4c03149.


Toward Next-Generation Molecular Imaging with a Clinical Low-Field (0.064 T) Point-of-Care MRI Scanner.

Iqbal N, Brittin D, Daluwathumullagamage P, Alam M, Senanayake I, Gafar A Anal Chem. 2024; 96(25):10348-10355.

PMID: 38857182 PMC: 11416731. DOI: 10.1021/acs.analchem.4c01299.


Molecules, Up Your Spins!.

Barskiy D Molecules. 2024; 29(8).

PMID: 38675641 PMC: 11052189. DOI: 10.3390/molecules29081821.


Spin Hyperpolarization in Modern Magnetic Resonance.

Eills J, Budker D, Cavagnero S, Chekmenev E, Elliott S, Jannin S Chem Rev. 2023; 123(4):1417-1551.

PMID: 36701528 PMC: 9951229. DOI: 10.1021/acs.chemrev.2c00534.


Proton-Only Sensing of Hyperpolarized [1,2-C]Pyruvate.

Mandzhieva I, Adelabu I, Chekmenev E, Theis T ACS Sens. 2022; 7(12):3773-3781.

PMID: 36414238 PMC: 10284333. DOI: 10.1021/acssensors.2c01608.


References
1.
Kennedy B, Kettunen M, Hu D, Brindle K . Probing lactate dehydrogenase activity in tumors by measuring hydrogen/deuterium exchange in hyperpolarized l-[1-(13)C,U-(2)H]lactate. J Am Chem Soc. 2012; 134(10):4969-77. PMC: 3303201. DOI: 10.1021/ja300222e. View

2.
Nikolaou P, Coffey A, Walkup L, Gust B, Whiting N, Newton H . Near-unity nuclear polarization with an open-source 129Xe hyperpolarizer for NMR and MRI. Proc Natl Acad Sci U S A. 2013; 110(35):14150-5. PMC: 3761567. DOI: 10.1073/pnas.1306586110. View

3.
Kettunen M, Kennedy B, Hu D, Brindle K . Spin echo measurements of the extravasation and tumor cell uptake of hyperpolarized [1-(13) C]lactate and [1-(13) C]pyruvate. Magn Reson Med. 2013; 70(5):1200-9. DOI: 10.1002/mrm.24591. View

4.
Golman K, Petersson J . Metabolic imaging and other applications of hyperpolarized 13C1. Acad Radiol. 2006; 13(8):932-42. DOI: 10.1016/j.acra.2006.06.001. View

5.
Sharma U, Baek H, Su M, Jagannathan N . In vivo 1H MRS in the assessment of the therapeutic response of breast cancer patients. NMR Biomed. 2011; 24(6):700-11. PMC: 4226268. DOI: 10.1002/nbm.1654. View