Evaluation of Ferromagnetism and Magnetic Resonance Imaging Artifacts of the Strecker Tantalum Vascular Stent

Overview
Date 1989 May 1
PMID 2507145
Citations 3
Authors
Affiliations
Soon will be listed here.
Abstract

A woven tantalum vascular stent (Strecker stent) was tested for ferromagnetism at 4.7 T and underwent magnetic resonance imaging (MRI) in vitro and in vivo at 1.5 T to evaluate the production of magnetic susceptibility artifacts. No ferromagnetism was detected. Spin-echo, phase reconstruction, and gradient echo images revealed a low level of susceptibility artifacts both in vitro and in vivo. Our findings demonstrate the feasibility of using MRI to evaluate blood vessels noninvasively following tantalum stent placement.

Citing Articles

In vitro image characteristics of an abdominal aortic stent graft: CTA versus 3D MRA.

Hilfiker P, Quick H, Schmidt M, Debatin J MAGMA. 1999; 8(1):27-32.

PMID: 10383090 DOI: 10.1007/BF02590632.


Magnetic resonance angiography of intravascular endoprostheses: investigation of three devices.

Laissy J, Grand C, Matos C, Struyven J, Berger J, Schouman-Claeys E Cardiovasc Intervent Radiol. 1995; 18(6):360-6.

PMID: 8591621 DOI: 10.1007/BF00338302.


Ferromagnetism and magnetic resonance artifacts of platinum embolization microcoils.

Marshall M, Teitelbaum G, Kim H, Deveikis J Cardiovasc Intervent Radiol. 1991; 14(3):163-6.

PMID: 1878906 DOI: 10.1007/BF02577720.

References
1.
NEW P, Rosen B, Brady T, Buonanno F, Kistler J, Burt C . Potential hazards and artifacts of ferromagnetic and nonferromagnetic surgical and dental materials and devices in nuclear magnetic resonance imaging. Radiology. 1983; 147(1):139-48. DOI: 10.1148/radiology.147.1.6828719. View

2.
Palmaz J, Windeler S, Garcia F, Tio F, Sibbitt R, REUTER S . Atherosclerotic rabbit aortas: expandable intraluminal grafting. Radiology. 1986; 160(3):723-6. DOI: 10.1148/radiology.160.3.2942964. View

3.
Teitelbaum G, Ortega H, Vinitski S, Stern H, Tsuruda J, Mitchell D . Low-artifact intravascular devices: MR imaging evaluation. Radiology. 1988; 168(3):713-9. DOI: 10.1148/radiology.168.3.3406402. View

4.
Liebman C, Messersmith R, Levin D, Lu C . MR imaging of inferior vena caval filters: safety and artifacts. AJR Am J Roentgenol. 1988; 150(5):1174-6. DOI: 10.2214/ajr.150.5.1174. View

5.
Shellock F, Crues J . High-field-strength MR imaging and metallic biomedical implants: an ex vivo evaluation of deflection forces. AJR Am J Roentgenol. 1988; 151(2):389-92. DOI: 10.2214/ajr.151.2.389. View