» Articles » PMID: 24930127

Woven Endobridge (WEB) Device for Endovascular Treatment of Ruptured Intracranial Wide-neck Aneurysms: a Single-center Experience

Overview
Journal Neuroradiology
Specialties Neurology
Radiology
Date 2014 Jun 16
PMID 24930127
Citations 34
Authors
Affiliations
Soon will be listed here.
Abstract

Introduction: Endovascular treatment of ruptured wide-neck bifurcation aneurysms presents a challenge. While still under evaluation, the Woven Endobridge (WEB) aneurysm embolization system has so far shown promising results in the treatment of complex bifurcation aneurysms. We aimed to evaluate the feasibility and short-term follow-up of endovascular treatment of ruptured wide-neck aneurysm with the WEB device.

Methods: Six patients referred to our institution for acute symptomatic subarachnoid hemorrhage (SAH) and treated with the WEB device were enrolled in this study. Clinical presentations, technical details, intraoperative and postoperative complications, and outcomes were recorded. Immediate and 3-month angiographic results were also evaluated.

Results: Three middle cerebral artery (MCA) and three anterior communicating artery aneurysms were treated between 1 and 14 days after rupturing. Average dome width was 5.8 mm (range 5-7), average neck size was 4.5 mm (range 4-5), and average dome-to-neck ratio was 1.3 (range 1-1.7). The WEB system was the exclusive treatment and was successfully deployed in all cases. Per procedural thromboembolic events occurred in two cases and were treated with intra-arterial administration of antiplatelet agents without any clinical consequences. The modified Rankin Scale (mRS) score at discharge was 0 for all patients. The 3-month angiographic follow-up showed adequate occlusion in four of our six patients (67 %).

Conclusion: From this preliminary study, the high feasibility rate and lack of need for systematic antiplatelet agents favor the WEB device providing a solution for endovascular treatment of ruptured wide-neck bifurcation aneurysms during the acute phase. However, further studies are needed to evaluate the complication rate and long-term efficiency.

Citing Articles

CLinical Assessment of WEB device in Ruptured aneurYSms (CLARYS): 12-month angiographic results of a multicenter study.

Spelle L, Herbreteau D, Caroff J, Barreau X, Ferre J, Fiehler J J Neurointerv Surg. 2022; 15(7):650-654.

PMID: 35882550 PMC: 10314057. DOI: 10.1136/neurintsurg-2022-018749.


Shape Memory Polymer-Based Endovascular Devices: Design Criteria and Future Perspective.

Pineda-Castillo S, Stiles A, Bohnstedt B, Lee H, Liu Y, Lee C Polymers (Basel). 2022; 14(13).

PMID: 35808573 PMC: 9269599. DOI: 10.3390/polym14132526.


CLinical Assessment of WEB device in Ruptured aneurYSms (CLARYS): results of 1-month and 1-year assessment of rebleeding protection and clinical safety in a multicenter study.

Spelle L, Herbreteau D, Caroff J, Barreau X, Ferre J, Fiehler J J Neurointerv Surg. 2021; 14(8):807-814.

PMID: 34493578 PMC: 9304095. DOI: 10.1136/neurintsurg-2021-017416.


Woven EndoBridge device for the treatment of ruptured intracranial aneurysms: A systematic review of clinical and angiographic results.

Xie Y, Tian H, Xiang B, Liu J, Xiang H Interv Neuroradiol. 2021; 28(2):240-249.

PMID: 34159820 PMC: 9131500. DOI: 10.1177/15910199211026712.


Safety and Efficacy of the Woven EndoBridge Device for Treatment of Ruptured Intracranial Aneurysms: A Systematic Review and Meta-analysis.

Essibayi M, Lanzino G, Brinjikji W AJNR Am J Neuroradiol. 2021; 42(9):1627-1632.

PMID: 34117016 PMC: 8423055. DOI: 10.3174/ajnr.A7174.


References
1.
Lubicz B, Klisch J, Gauvrit J, Szikora I, Leonardi M, Liebig T . WEB-DL endovascular treatment of wide-neck bifurcation aneurysms: short- and midterm results in a European study. AJNR Am J Neuroradiol. 2014; 35(3):432-8. PMC: 7964738. DOI: 10.3174/ajnr.A3869. View

2.
Cognard C, Pierot L, Anxionnat R, Ricolfi F . Results of embolization used as the first treatment choice in a consecutive nonselected population of ruptured aneurysms: clinical results of the Clarity GDC study. Neurosurgery. 2011; 69(4):837-41. DOI: 10.1227/NEU.0b013e3182257b30. View

3.
Pierot L, Cognard C, Anxionnat R, Ricolfi F . Ruptured intracranial aneurysms: factors affecting the rate and outcome of endovascular treatment complications in a series of 782 patients (CLARITY study). Radiology. 2010; 256(3):916-23. DOI: 10.1148/radiol.10092209. View

4.
Ding Y, Lewis D, Kadirvel R, Dai D, Kallmes D . The Woven EndoBridge: a new aneurysm occlusion device. AJNR Am J Neuroradiol. 2011; 32(3):607-11. PMC: 8013102. DOI: 10.3174/ajnr.A2399. View

5.
Moret J, Cognard C, Weill A, Castaings L, Rey A . [Reconstruction technic in the treatment of wide-neck intracranial aneurysms. Long-term angiographic and clinical results. Apropos of 56 cases]. J Neuroradiol. 1997; 24(1):30-44. View