» Articles » PMID: 24917454

Intermittent Aerosol Delivery to the Lungs During High-flow Nasal Cannula Therapy

Overview
Journal Respir Care
Publisher Mary Ann Liebert
Specialty Pulmonary Medicine
Date 2014 Jun 12
PMID 24917454
Citations 24
Authors
Affiliations
Soon will be listed here.
Abstract

Introduction: Use of submicrometer particles combined with condensational growth techniques has been proposed to reduce drug losses within components of high-flow nasal cannula therapy systems and to enhance the dose reaching the lower respiratory tract. These methods have been evaluated using continuous inhalation flow rather than realistic inhalation/exhalation breathing cycles. The goal of this study was to evaluate in vitro aerosol drug delivery using condensational growth techniques during high-flow nasal cannula therapy using realistic breathing profiles and incorporating intermittent aerosol delivery techniques.

Methods: A mixer-heater combined with a vibrating mesh nebulizer was used to generate a submicrometer aerosol using a formulation of 0.2% albuterol sulfate and 0.2% sodium chloride in water. Delivery efficiency of the aerosol for 1 min through a nasal cannula was considered using an intermittent delivery regime with aerosol being emitted for either the entire inhalation time (2 s) or half of the inhalation period (1 s) and compared with continuous delivery. The deposition of the aerosol was evaluated in the nasal delivery components (ventilator tubing and cannula) and an in vitro adult nose-mouth-throat (NMT) model using 3 realistic breathing profiles.

Results: Significant improvements in dose delivered to the exit of the NMT model (ex-NMT) were observed for both condensational growth methods using intermittent aerosol delivery compared with continuous delivery, and increasing the tidal volume was found useful. The combination of the largest tidal volume with the shortest intermittent delivery time resulted in the lowest respiration losses and the highest ex-NMT delivered dose.

Conclusions: Intermittent aerosol delivery using realistic breathing profiles of submicrometer condensational growth aerosols was found to be efficient in delivering nasally administered drugs in an in vitro airway model.

Citing Articles

Performance Characterisation of the Airvo2 Nebuliser Adapter in Combination with the Aerogen Solo Vibrating Mesh Nebuliser for in Line Aerosol Therapy during High Flow Nasal Oxygen Therapy.

MacLoughlin R, Mac Giolla Eain M Pharmaceutics. 2024; 16(4).

PMID: 38675226 PMC: 11053618. DOI: 10.3390/pharmaceutics16040565.


In-Line Aerosol Therapy via Nasal Cannula during Adult and Paediatric Normal, Obstructive, and Restrictive Breathing.

Mac Giolla Eain M, MacLoughlin R Pharmaceutics. 2023; 15(12).

PMID: 38140020 PMC: 10747070. DOI: 10.3390/pharmaceutics15122679.


Impact of high- and low-flow nebulised saline on airway hydration and mucociliary transport.

Kelly S, Valentine M, Chua W, Tatkov S ERJ Open Res. 2023; 9(3).

PMID: 37313398 PMC: 10258716. DOI: 10.1183/23120541.00724-2022.


In Vitro Model for Analysis of High-Flow Aerosol Delivery During Continuous Nebulization.

McPeck M, Moon J, Jayakumaran J, Smaldone G Respir Care. 2023; 68(9):1213-1220.

PMID: 37253606 PMC: 10468165. DOI: 10.4187/respcare.10643.


Aerosol delivery through high-flow nasal therapy: Technical issues and clinical benefits.

Calabrese C, Annunziata A, Mariniello D, Allocca V, Imitazione P, Cauteruccio R Front Med (Lausanne). 2023; 9:1098427.

PMID: 36743674 PMC: 9889634. DOI: 10.3389/fmed.2022.1098427.