MRI-conditional Pacemakers: Current Perspectives

Overview
Publisher Dove Medical Press
Date 2014 May 23
PMID 24851058
Citations 19
Authors
Affiliations
Soon will be listed here.
Abstract

Use of both magnetic resonance imaging (MRI) and pacing devices has undergone remarkable growth in recent years, and it is estimated that the majority of patients with pacemakers will need an MRI during their lifetime. These investigations will generally be denied due to the potentially dangerous interactions between cardiac devices and the magnetic fields and radio frequency energy used in MRI. Despite the increasing reports of uneventful scanning in selected patients with conventional pacemakers under close surveillance, MRI is still contraindicated in those circumstances and cannot be considered a routine procedure. These limitations prompted a series of modifications in generator and lead engineering, designed to minimize interactions that could compromise device function and patient safety. The resulting MRI-conditional pacemakers were first introduced in 2008 and the clinical experience gathered so far supports their safety in the MRI environment if certain conditions are fulfilled. With this technology, new questions and controversies arise regarding patient selection, clinical impact, and cost-effectiveness. In this review, we discuss the potential risks of MRI in patients with electronic cardiac devices and present updated information regarding the features of MRI-conditional pacemakers and the clinical experience with currently available models. Finally, we provide some guidance on how to scan patients who have these devices and discuss future directions in the field.

Citing Articles

Latest Developments in Adapting Deep Learning for Assessing TAVR Procedures and Outcomes.

Tahir A, Mutlu O, Bensaali F, Ward R, Ghareeb A, Helmy S J Clin Med. 2023; 12(14).

PMID: 37510889 PMC: 10381346. DOI: 10.3390/jcm12144774.


MRI of Patients with Cardiac Implantable Electronic Devices.

Martinez J, Ennis D Curr Cardiovasc Imaging Rep. 2022; 12(7).

PMID: 36117853 PMC: 9477432. DOI: 10.1007/s12410-019-9502-8.


Cardiac Implantable Electronic Devices and Consumer Electronic Devices: The Proof Is in the Front Pocket.

Ellis C, King N J Innov Card Rhythm Manag. 2022; 13(7):5073-5076.

PMID: 35949651 PMC: 9359421. DOI: 10.19102/icrm.2022.130706.


Lumbar Intervertebral Disc Herniation: Annular Closure Devices and Key Design Requirements.

Guardado A, Baker A, Weightman A, Hoyland J, Cooper G Bioengineering (Basel). 2022; 9(2).

PMID: 35200401 PMC: 8869316. DOI: 10.3390/bioengineering9020047.


Should everyone have an MRI in heart failure?.

Nikolaidou C, Karamitsos T Cardiovasc Diagn Ther. 2020; 10(3):549-553.

PMID: 32695635 PMC: 7369283. DOI: 10.21037/cdt.2019.12.06.


References
1.
Greenspon A, Patel J, Lau E, Ochoa J, Frisch D, Ho R . Trends in permanent pacemaker implantation in the United States from 1993 to 2009: increasing complexity of patients and procedures. J Am Coll Cardiol. 2012; 60(16):1540-5. DOI: 10.1016/j.jacc.2012.07.017. View

2.
Zikria J, Machnicki S, Rhim E, Bhatti T, Graham R . MRI of patients with cardiac pacemakers: a review of the medical literature. AJR Am J Roentgenol. 2011; 196(2):390-401. DOI: 10.2214/AJR.10.4239. View

3.
Wollmann C, Steiner E, Vock P, Ndikung B, Mayr H . Monocenter feasibility study of the MRI compatibility of the Evia pacemaker in combination with Safio S pacemaker lead. J Cardiovasc Magn Reson. 2012; 14:67. PMC: 3482396. DOI: 10.1186/1532-429X-14-67. View

4.
Pulver A, Puchalski M, Bradley D, Minich L, Su J, Saarel E . Safety and imaging quality of MRI in pediatric and adult congenital heart disease patients with pacemakers. Pacing Clin Electrophysiol. 2009; 32(4):450-6. DOI: 10.1111/j.1540-8159.2009.02304.x. View

5.
Nazarian S, Hansford R, Roguin A, Goldsher D, Zviman M, Lardo A . A prospective evaluation of a protocol for magnetic resonance imaging of patients with implanted cardiac devices. Ann Intern Med. 2011; 155(7):415-24. PMC: 4337840. DOI: 10.7326/0003-4819-155-7-201110040-00004. View