Pacemaker Lead Tip Heating in Abandoned and Pacemaker-attached Leads at 1.5 Tesla MRI

Overview
Date 2011 Jan 29
PMID 21274985
Citations 51
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: To assess the risk of RF-induced heating in pacemaker-attached and abandoned leads using in vitro temperature measurements at 1.5 Tesla as a function of lead length.

Materials And Methods: Five custom lead lengths, 20-60 cm, were exposed to a uniform magnitude and phase radiofrequency electric field to examine the effect of lead length on pacemaker lead tip heating for pacemaker-attached and abandoned pacemaker leads.

Results: Abandoned and pacemaker-attached leads show resonant heating behavior and maximum heating occurs at different lead lengths due to the differences in termination conditions. For clinical lead lengths (40-60 cm) abandoned leads exhibited greater lead tip heating compared with pacemaker-attached leads.

Conclusion: Current recommendations for MRI pacemaker safety should highlight the possible increased risk for patients with abandoned leads as compared to pacemaker-attached leads.

Citing Articles

MRI in Patients with Cardiovascular Implantable Electronic Devices and Fractured or Abandoned Leads.

Greenhill M, Rangan P, Su W, Weiss J, Zawaneh M, Unzek S Radiol Cardiothorac Imaging. 2024; 6(3):e230303.

PMID: 38869431 PMC: 11211945. DOI: 10.1148/ryct.230303.


Modeling and measurement of lead tip heating and resonant length for implanted, insulated wires.

Bardwell Speltz L, Lee S, Shu Y, Tarasek M, Trzasko J, Foo T Magn Reson Med. 2024; 92(4):1714-1727.

PMID: 38818673 PMC: 11414523. DOI: 10.1002/mrm.30145.


How we do it: Cardiac implantable devices are not a contraindication to MRI: time for a paradigm shift.

De Leon-Benedetti L, Ramirez-Suarez K, Otero H, Rapp J, Biko D, Smith C Pediatr Radiol. 2024; 54(6):863-875.

PMID: 38488925 DOI: 10.1007/s00247-024-05902-y.


Current Insights and Novel Cardiovascular Magnetic Resonance-Based Techniques in the Prognosis of Non-Ischemic Dilated Cardiomyopathy.

Perone F, Dentamaro I, La Mura L, Alifragki A, Marketou M, Cavarretta E J Clin Med. 2024; 13(4).

PMID: 38398330 PMC: 10889760. DOI: 10.3390/jcm13041017.


SCMR expert consensus statement for cardiovascular magnetic resonance of patients with a cardiac implantable electronic device.

Kim D, Collins J, White J, Hanneman K, Lee D, Patel A J Cardiovasc Magn Reson. 2024; 26(1):100995.

PMID: 38219955 PMC: 11211236. DOI: 10.1016/j.jocmr.2024.100995.