» Articles » PMID: 248270

Methicillin-resistant Staphylococcus Aureus Susceptibility Testing by an Automated System, Autobac I

Overview
Specialty Pharmacology
Date 1978 May 1
PMID 248270
Citations 19
Authors
Affiliations
Soon will be listed here.
Abstract

The Autobac I system was used to evaluate the antibiotic susceptibility pattern of methicillin-resistant Staphylococcus aureus isolates. The results of the Autobac I were compared with the results of the disk diffusion method. The disk diffusion susceptibility pattern showed resistance to methicillin/oxacillin, penicillin, erythromycin, clindamycin, and kanamycin. All isolates were susceptible to cephalothin, chloramphenicol, tetracycline, and gentamicin. There was at least 96% agreement using the Autobac I system with all antibiotics except methicillin and clindamycin. Seventy percent of 57 isolates were recorded as susceptible to methicillin, whereas 9% had an intermediate susceptibility. With clindamycin, 14% were recorded as susceptible and 7% were recorded as intermediate. Upon prolonged incubation of the Autobac I cuvette, the agreement between the two methods was 44% for methicillin and 93% for clindamycin. Changes in the environmental conditions, such as use of 5% sodium chloride broth and a 32 degrees C incubation temperature, did not increase the detection of methicillin-resistant isolates by the Autobac I system.

Citing Articles

Evaluation of the Cathra inoculating device for susceptibility testing of methicillin-susceptible and methicillin-resistant Staphylococcus aureus strains.

Boyce J, Bonner M, LOCKWOOD W Antimicrob Agents Chemother. 1982; 22(5):901-2.

PMID: 6924824 PMC: 185680. DOI: 10.1128/AAC.22.5.901.


Susceptibility testing of multidrug-resistant Staphylococcus aureus with the Sceptor microdilution system.

Blair H, Cleary T J Clin Microbiol. 1983; 18(1):194-6.

PMID: 6885987 PMC: 270767. DOI: 10.1128/jcm.18.1.194-196.1983.


Examination of Pseudomonas aeruginosa-gentamicin discrepancies encountered in an Autobac I-disk diffusion comparison.

Mayo J, Kiehn T, Wong B, Bernard E, Armstrong D Antimicrob Agents Chemother. 1982; 21(3):412-5.

PMID: 6808909 PMC: 181906. DOI: 10.1128/AAC.21.3.412.


Successful use of broth microdilution in susceptibility tests for methicillin-resistant (heteroresistant) staphylococci.

Thornsberry C, McDougal L J Clin Microbiol. 1983; 18(5):1084-91.

PMID: 6643661 PMC: 272846. DOI: 10.1128/jcm.18.5.1084-1091.1983.


Interlaboratory variation of antibiograms of methicillin-resistant and methicillin-susceptible Staphylococcus aureus strains with conventional and commercial testing systems.

Aldridge K, JANNEY A, Sanders C, Marier R J Clin Microbiol. 1983; 18(5):1226-36.

PMID: 6605976 PMC: 272871. DOI: 10.1128/jcm.18.5.1226-1236.1983.


References
1.
Barry A, Badal R . Reliability of the microdilution technic for detection of methicillin-resistant strains of staphylococcus aureus. Am J Clin Pathol. 1977; 67(5):489-95. DOI: 10.1093/ajcp/67.5.489. View

2.
Hewitt J, COE A, PARKER M . The detection of methicillin resistance in Staphylococcus aureus. J Med Microbiol. 1969; 2(4):443-56. DOI: 10.1099/00222615-2-4-443. View

3.
BAUER A, KIRBY W, SHERRIS J, Turck M . Antibiotic susceptibility testing by a standardized single disk method. Am J Clin Pathol. 1966; 45(4):493-6. View

4.
Barrett F, McGehee Jr R, FINLAND M . Methicillin-resistant Staphylococcus aureus at Boston City Hospital. Bacteriologic and epidemiologic observations. N Engl J Med. 1968; 279(9):441-8. DOI: 10.1056/NEJM196808292790901. View

5.
PLORDE J, SHERRIS J . Staphylococcal resistance to antibiotics: origin, measurement, and epidemiology. Ann N Y Acad Sci. 1974; 236(0):413-34. DOI: 10.1111/j.1749-6632.1974.tb41507.x. View