Study of Electrode Pattern Design for a CZT-based PET Detector

Overview
Journal Phys Med Biol
Publisher IOP Publishing
Date 2014 May 3
PMID 24786208
Citations 10
Authors
Affiliations
Soon will be listed here.
Abstract

We are developing a 1 mm resolution small animal positron emission tomography (PET) system using 3D positioning cadmium zinc telluride photon detectors comprising 40 mm × 40 mm × 5 mm crystals metalized with a cross-strip electrode pattern with a 1 mm anode strip pitch. We optimized the electrode pattern design for intrinsic sensitivity and spatial, energy and time resolution performance using a test detector comprising cathode and steering electrode strips of varying dimensions. The study found 3 and 5 mm width cathode strips locate charge-shared photon interactions near cathode strip boundaries with equal precision. 3 mm width cathode strips exhibited large time resolution variability as a function of photon interaction location between the anode and cathode planes (~26 to ~127.5 ns full width at half maximum (FWHM) for 0.5 mm and 4.2 mm depths, respectively). 5 mm width cathode strips by contrast exhibited more stable time resolution for the same interaction locations (~34 to ~83 ns FWHM), provided more linear spatial positioning in the direction orthogonal to the electrode planes, and as much as 68.4% improvement in photon sensitivity over the 3 mm wide cathode strips. The results were understood by analyzing the cathode strips' weighting functions, which indicated a stronger 'small pixel' effect in the 3 mm wide cathode strips. Photon sensitivity and anode energy resolution were seen to improve with decreasing steering electrode bias from 0 to -80 V w.r.t. the anode potential. A slight improvement in energy resolution was seen for wider steering electrode strips (400 versus 100 µm) for charge-shared photon interactions. Although this study successfully focused on electrode pattern features for PET performance, the results are generally applicable to semiconductor photon detectors employing cross-trip electrode patterns.

Citing Articles

Performance Investigations of Two Channel Readout Configurations on the Cross-Strip Cadmium Zinc Telluride Detector.

Enlow E, Wang Y, Shoop G, Abbaszadeh S IEEE Trans Radiat Plasma Med Sci. 2024; 8(8):886-892.

PMID: 39534386 PMC: 11556448. DOI: 10.1109/trpms.2024.3411522.


Quantum Entanglement Filtering: A PET feasibility study in imaging dual-positron and prompt gamma emission via Monte Carlo simulation.

Romanchek G, Shoop G, Gholami K, Enlow E, Abbaszadeh S IEEE Trans Radiat Plasma Med Sci. 2024; 8(8):916-925.

PMID: 39507126 PMC: 11540415. DOI: 10.1109/trpms.2024.3388872.


Reduction of the Low Energy Threshold through New Modular Data Acquisition Electronics for Cross-strip Cadmium Zinc Telluride (CZT) based PET System.

Wang Y, Abbaszadeh S IEEE Nucl Sci Symp Conf Rec (1997). 2024; 2021.

PMID: 39469219 PMC: 11517392. DOI: 10.1109/nss/mic44867.2021.9875680.


Radiation Detectors and Sensors in Medical Imaging.

Michail C, Liaparinos P, Kalyvas N, Kandarakis I, Fountos G, Valais I Sensors (Basel). 2024; 24(19).

PMID: 39409289 PMC: 11478476. DOI: 10.3390/s24196251.


Impact of Flexible Circuit Bonding and System Integration on Energy Resolution of Cross-Strip CZT Detectors.

Enlow E, Diba M, Clayton J, Harris B, Abbaszadeh S IEEE Trans Radiat Plasma Med Sci. 2024; 7(6):580-586.

PMID: 38468608 PMC: 10927013. DOI: 10.1109/trpms.2023.3256406.


References
1.
Gu Y, Matteson J, Skelton R, Deal A, Stephan E, Duttweiler F . Study of a high-resolution, 3D positioning cadmium zinc telluride detector for PET. Phys Med Biol. 2011; 56(6):1563-84. PMC: 3274175. DOI: 10.1088/0031-9155/56/6/004. View

2.
Meng L, He Z . Exploring the limiting timing resolution for large volume CZT detectors with waveform analysis. IEEE Trans Nucl Sci. 2017; 550(1-2):435-445. PMC: 5333786. DOI: 10.1016/j.nima.2005.04.076. View

3.
Levin C, Foudray A, Habte F . Impact of high energy resolution detectors on the performance of a PET system dedicated to breast cancer imaging. Phys Med. 2007; 21 Suppl 1:28-34. DOI: 10.1016/S1120-1797(06)80020-7. View

4.
Gu Y, Pratx G, Lau F, Levin C . Effects of multiple-interaction photon events in a high-resolution PET system that uses 3-D positioning detectors. Med Phys. 2010; 37(10):5494-508. PMC: 2962664. DOI: 10.1118/1.3483262. View

5.
Levin C . Promising new photon detection concepts for high-resolution clinical and preclinical PET. J Nucl Med. 2012; 53(2):167-70. PMC: 3674960. DOI: 10.2967/jnumed.110.084343. View