Background: The calculation of arterial oxygen saturation (SpO2) relies heavily on the amplitude information of the high-quality photoplethysmographic (PPG) signals, which could be contaminated by motion artifacts (MA) during monitoring.

Methods: A new method combining temporally constrained independent component analysis (cICA) and adaptive filters is presented here to extract the clean PPG signals from the MA corrupted PPG signals with the amplitude information reserved. The underlying PPG signal could be extracted from the MA contaminated PPG signals automatically by using cICA algorithm. Then the amplitude information of the PPG signals could be recovered by using adaptive filters.

Results: Compared with conventional ICA algorithms, the proposed approach is permutation and scale ambiguity-free. Numerical examples with both synthetic datasets and real-world MA corrupted PPG signals demonstrate that the proposed method could remove the MA from MA contaminated PPG signals more effectively than the two existing FFT-LMS and moving average filter (MAF) methods.

Conclusions: This paper presents a new method which combines the cICA algorithm and adaptive filter to extract the underlying PPG signals from the MA contaminated PPG signals with the amplitude information reserved. The new method could be used in the situations where one wants to extract the interested source automatically from the mixed observed signals with the amplitude information reserved. The results of study demonstrated the efficacy of this proposed method.

Citing Articles

WF-PPG: A Wrist-finger Dual-Channel Dataset for Studying the Impact of Contact Pressure on PPG Morphology.

Ho M, Pham H, Saeed A, Ma D Sci Data. 2025; 12(1):200.

PMID: 39900957 PMC: 11790827. DOI: 10.1038/s41597-025-04453-7.


On the development of low power wearable devices for assessment of physiological vital parameters: a systematic review.

Jegan R, Nimi W Z Gesundh Wiss. 2023; :1-16.

PMID: 37361281 PMC: 10068243. DOI: 10.1007/s10389-023-01893-6.


New pulse oximetry detection based on the light absorbance ratio as determined from amplitude modulation indexes in the time and frequency domains.

Kainan P, Sinchai A, Tuwanut P, Wardkein P Biomed Signal Process Control. 2022; 75:103627.

PMID: 36267662 PMC: 9558448. DOI: 10.1016/j.bspc.2022.103627.


Wrist Photoplethysmography Signal Quality Assessment for Reliable Heart Rate Estimate and Morphological Analysis.

Moscato S, Giudice S, Massaro G, Chiari L Sensors (Basel). 2022; 22(15).

PMID: 35957395 PMC: 9370973. DOI: 10.3390/s22155831.


Photoplethysmogram Recording Length: Defining Minimal Length Requirement from Dynamical Characteristics.

Sviridova N, Zhao T, Nakano A, Ikeguchi T Sensors (Basel). 2022; 22(14).

PMID: 35890834 PMC: 9324273. DOI: 10.3390/s22145154.


References
1.
Goldman J, Petterson M, Kopotic R, Barker S . Masimo signal extraction pulse oximetry. J Clin Monit Comput. 2003; 16(7):475-83. DOI: 10.1023/a:1011493521730. View

2.
Yousefi R, Nourani M, Panahi I . Adaptive cancellation of motion artifact in wearable biosensors. Annu Int Conf IEEE Eng Med Biol Soc. 2013; 2012:2004-8. DOI: 10.1109/EMBC.2012.6346350. View

3.
Bell A, Sejnowski T . An information-maximization approach to blind separation and blind deconvolution. Neural Comput. 1995; 7(6):1129-59. DOI: 10.1162/neco.1995.7.6.1129. View

4.
Poh M, Swenson N, Picard R . Motion-tolerant magnetic earring sensor and wireless earpiece for wearable photoplethysmography. IEEE Trans Inf Technol Biomed. 2010; 14(3):786-94. DOI: 10.1109/TITB.2010.2042607. View

5.
Asada H, Shaltis P, Reisner A, Rhee S, Hutchinson R . Mobile monitoring with wearable photoplethysmographic biosensors. IEEE Eng Med Biol Mag. 2003; 22(3):28-40. DOI: 10.1109/memb.2003.1213624. View