» Articles » PMID: 24550367

Improvement in Outflow Facility by Two Novel Microinvasive Glaucoma Surgery Implants

Overview
Specialty Ophthalmology
Date 2014 Feb 20
PMID 24550367
Citations 14
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: To determine improvement in outflow facility (C) in human anterior segments implanted with a novel Schlemm's canal scaffold or two trabecular micro-bypasses.

Methods: Human anterior segments were isolated from 12 pairs of eyes from donors with no history of ocular disease and then perfused at 50, 40, 30, 20, and 10 mm Hg pressures for 10 minutes each. Baseline C was calculated from perfusion pressures and flow rates. The scaffold was implanted into Schlemm's canal of one anterior segment, and two micro-bypasses were implanted three clock-hours apart in the contralateral anterior segment. Outflow facility and resistance were compared at various standardized perfusion pressures and between each device.

Results: Compared to baseline, C increased by 0.16 ± 0.12 μL/min/mm Hg (74%) with the scaffold, and 0.08 ± 0.12 μL/min/mm Hg (34%) with two micro-bypasses. The scaffold increased C at perfusion pressures of 50, 40, 30, and 20 mm Hg (P < 0.005). Two micro-bypasses increased C at a perfusion pressure of 40 mm Hg (P < 0.05).

Conclusions: Both implants effectively increased C in human eyes ex vivo. The scaffold increased C by a greater percentage (73% vs. 34%) and at a greater range of perfusion pressures (20 to 50 mm Hg vs. 40 mm Hg) than the two micro-bypasses, suggesting that the 8-mm dilation of Schlemm's canal by the scaffold may have additional benefits in lowering the outflow resistance. The Hydrus Microstent scaffold may be an effective therapy for increasing outflow facility and thus reducing the IOP in patients with glaucoma.

Citing Articles

Minimally invasive glaucoma surgery: comparison of Hydrus microstent with iStent in primary open-angle glaucoma.

Komzak K, Allen P, Toh T BMJ Open Ophthalmol. 2025; 10(1).

PMID: 39947711 PMC: 11831292. DOI: 10.1136/bmjophth-2024-001946.


Analyzing the Shortcomings of Trabecular Micro-bypass Stents for Surgical Management of Glaucoma.

Dada T, Beri N J Curr Glaucoma Pract. 2024; 18(1):1-3.

PMID: 38585166 PMC: 10997962. DOI: 10.5005/jp-journals-10078-1439.


Spotlight on Schlemm's Canal MicroStent Injection in Patients with Glaucoma.

Wagner I, Ang B, Checo L, Simsek D, Draper C, Dorairaj S Clin Ophthalmol. 2023; 17:1557-1564.

PMID: 37288002 PMC: 10243342. DOI: 10.2147/OPTH.S388293.


Comparison of Safety and Efficacy of Hydrus and iStent Combined with Phacoemulsyfication in Open Angle Glaucoma Patients: 24-Month Follow-Up.

Jablonska J, Lewczuk K, Rekas M Int J Environ Res Public Health. 2023; 20(5).

PMID: 36901162 PMC: 10002347. DOI: 10.3390/ijerph20054152.


Comparison of Hydrus and iStent microinvasive glaucoma surgery implants in combination with phacoemulsification for treatment of open-angle glaucoma: systematic review and network meta-analysis.

Hu R, Guo D, Hong N, Xuan X, Wang X BMJ Open. 2022; 12(6):e051496.

PMID: 35705355 PMC: 9204447. DOI: 10.1136/bmjopen-2021-051496.


References
1.
Moses R . Circumferential flow in Schlemm's canal. Am J Ophthalmol. 1979; 88(3 Pt 2):585-91. DOI: 10.1016/0002-9394(79)90519-1. View

2.
Quigley H . Glaucoma. Lancet. 2011; 377(9774):1367-77. DOI: 10.1016/S0140-6736(10)61423-7. View

3.
Saheb H, Ahmed I . Micro-invasive glaucoma surgery: current perspectives and future directions. Curr Opin Ophthalmol. 2012; 23(2):96-104. DOI: 10.1097/ICU.0b013e32834ff1e7. View

4.
Camras L, Yuan F, Fan S, Samuelson T, Ahmed I, Schieber A . A novel Schlemm's Canal scaffold increases outflow facility in a human anterior segment perfusion model. Invest Ophthalmol Vis Sci. 2012; 53(10):6115-21. DOI: 10.1167/iovs.12-9570. View

5.
Tsai J . Medication adherence in glaucoma: approaches for optimizing patient compliance. Curr Opin Ophthalmol. 2006; 17(2):190-5. DOI: 10.1097/01.icu.0000193078.47616.aa. View