» Articles » PMID: 24511401

Fracture of Two Moderately Cross-Linked Polyethylene Tibial Inserts in a TKR Patient

Overview
Journal Case Rep Orthop
Publisher Wiley
Specialty Orthopedics
Date 2014 Feb 11
PMID 24511401
Citations 2
Authors
Affiliations
Soon will be listed here.
Abstract

Highly cross-linked polyethylene has become the gold standard in total hip replacement for its wear resistance. Moderately crosslinked polyethylene is now available for total knee replacement (TKR), although concerns about reduced mechanical strength have prevented widespread adoption. The purpose of this report is to describe an unusual case where a patient underwent cruciate retaining TKR using a moderately crosslinked polyethylene tibial insert that went on to fracture twice in the same location across the primary and first revision surgery. The first tibial insert was 10 mm thick and was implanted for 16 months. The second tibial insert was 15 mm thick and was implanted for 11 months. Both fractured along the posterior aspect of the medial articular surface. The lack of a specific event leading to these fractures and the fact that they occurred twice in the same location in the same patient suggest that caution is still necessary regarding the introduction of crosslinked polyethylene for TKR surgery.

Citing Articles

Early failure of sequentially annealed polyethylene in total knee arthroplasty.

Sonn K, Meneghini R Arthroplast Today. 2020; 6(1):18-22.

PMID: 32211470 PMC: 7083714. DOI: 10.1016/j.artd.2019.12.005.


Influence of conformity on the wear of total knee replacement: An experimental study.

Brockett C, Carbone S, Fisher J, Jennings L Proc Inst Mech Eng H. 2017; 232(2):127-134.

PMID: 29251167 PMC: 5808816. DOI: 10.1177/0954411917746433.

References
1.
Lachiewicz P, Geyer M . The use of highly cross-linked polyethylene in total knee arthroplasty. J Am Acad Orthop Surg. 2011; 19(3):143-51. DOI: 10.5435/00124635-201103000-00003. View

2.
Teeter M, Yuan X, Naudie D, Holdsworth D . Technique to quantify subsurface cracks in retrieved polyethylene components using micro-CT. J Long Term Eff Med Implants. 2011; 20(1):27-34. DOI: 10.1615/jlongtermeffmedimplants.v20.i1.40. View

3.
Goldstein M, Ast M, DiMaio F . Acute posttraumatic catastrophic failure of a second-generation, highly cross-linked ultra-high-molecular-weight polyethylene patellar component. Orthopedics. 2012; 35(7):e1119-21. DOI: 10.3928/01477447-20120621-35. View

4.
Popoola O, Yao J, Johnson T, Blanchard C . Wear, delamination, and fatigue resistance of melt-annealed highly crosslinked UHMWPE cruciate-retaining knee inserts under activities of daily living. J Orthop Res. 2010; 28(9):1120-6. DOI: 10.1002/jor.21104. View

5.
Willie B, Foot L, Prall M, Bloebaum R . Surface damage analysis of retrieved highly crosslinked polyethylene tibial components after short-term implantation. J Biomed Mater Res B Appl Biomater. 2007; 85(1):114-24. DOI: 10.1002/jbm.b.30923. View