Abstract

The ideal treatment for unstable thoracolumbar fractures remains controversial with posterior reduction and stabilization, anterior reduction and stabilization, combined posterior and anterior reduction and stabilization, and even nonoperative management advocated. Short segment posterior osteosynthesis of these fractures has less comorbidities compared with the other operative approaches but settles into kyphosis over time. Biomechanical comparison of the divergent bridge construct versus the parallel tension band construct was performed for anteriorly destabilized T11-L1 spine segments using three different models: (1) finite element analysis (FEA), (2) a synthetic model, and (3) a human cadaveric model. Outcomes measured were construct stiffness and ultimate failure load. Our objective was to determine if the divergent pedicle screw bridge construct would provide more resistance to kyphotic deforming forces. All three modalities showed greater stiffness with the divergent bridge construct. The FEA calculated a stiffness of 21.6 N/m for the tension band construct versus 34.1 N/m for the divergent bridge construct. The synthetic model resulted in a mean stiffness of 17.3 N/m for parallel tension band versus 20.6 N/m for the divergent bridge (p = 0.03), whereas the cadaveric model had an average stiffness of 15.2 N/m in the parallel tension band compared with 18.4 N/m for the divergent bridge (p = 0.02). Ultimate failure load with the cadaveric model was found to be 622 N for the divergent bridge construct versus 419 N (p = 0.15) for the parallel tension band construct. This study confirms our clinical experience that the short posterior divergent bridge construct provides greater stiffness for the management of unstable thoracolumbar fractures.

Citing Articles

Numerical analysis of multi-level versus short instrumentation for the treatment of thoracolumbar fractures.

Hubner A, Gasparin D, de Meira Junior A, Israel C, Dambros J, Ribeiro M Eur J Orthop Surg Traumatol. 2015; 25 Suppl 1:S213-7.

PMID: 25733345 DOI: 10.1007/s00590-015-1612-7.

References
1.
Gaines Jr R, Carson W, Satterlee C, Groh G . Experimental evaluation of seven different spinal fracture internal fixation devices using nonfailure stability testing. The load-sharing and unstable-mechanism concepts. Spine (Phila Pa 1976). 1991; 16(8):902-9. DOI: 10.1097/00007632-199108000-00007. View

2.
McDonough P, Davis R, Tribus C, Zdeblick T . The management of acute thoracolumbar burst fractures with anterior corpectomy and Z-plate fixation. Spine (Phila Pa 1976). 2004; 29(17):1901-8. DOI: 10.1097/01.brs.0000137059.03557.1d. View

3.
Tezeren G, Kuru I . Posterior fixation of thoracolumbar burst fracture: short-segment pedicle fixation versus long-segment instrumentation. J Spinal Disord Tech. 2005; 18(6):485-8. DOI: 10.1097/01.bsd.0000149874.61397.38. View

4.
Parker J, Lane J, Karaikovic E, Gaines R . Successful short-segment instrumentation and fusion for thoracolumbar spine fractures: a consecutive 41/2-year series. Spine (Phila Pa 1976). 2000; 25(9):1157-70. DOI: 10.1097/00007632-200005010-00018. View

5.
McKinley T, McLain R, Yerby S, Sharkey N, Smith T . Characteristics of pedicle screw loading. Effect of surgical technique on intravertebral and intrapedicular bending moments. Spine (Phila Pa 1976). 1999; 24(1):18-24, discussion 25. DOI: 10.1097/00007632-199901010-00005. View