An Investigation of Gantry Angle Data Accuracy for Cine-mode EPID Images Acquired During Arc IMRT

Overview
Date 2014 Jan 16
PMID 24423849
Citations 7
Authors
Affiliations
Soon will be listed here.
Abstract

EPID images acquired in cine mode during arc therapy have inaccurate gantry angles recorded in their image headers. In this work, methods were developed to assess the accuracy of the gantry potentiometer for linear accelerators. As well, assessments of the accuracy of other, more accessible, sources of gantry angle information (i.e., treatment log files, analysis of EPID image headers) were investigated. The methods used in this study are generally applicable to any linear accelerator unit, and have been demonstrated here with Clinac/Trilogy systems. Gantry angle data were simultaneously acquired using three methods: i) a direct gantry potentiometer measurement, ii) an incremental rotary encoder, and iii) a custom-made radiographic gantry-angle phantom which produced unique wire intersections as a function of gantry angle. All methods were compared to gantry angle data from the EPID image header and the linac MLC DynaLog file. The encoder and gantry-angle phantom were used to validate the accuracy of the linac's potentiometer. The EPID image header gantry angles and the DynaLog file gantry angles were compared to the potentiometer. The encoder and gantry-angle phantom mean angle differences with the potentiometer were 0.13° ± 0.14° and 0.10°± 0.30°, respectively. The EPID image header angles analyzed in this study were within ± 1° of the potentiometer angles only 35% of the time. In some cases, EPID image header gantry angles disagreed by as much as 3° with the potentiometer. A time delay in frame acquisition was determined using the continuous acquisition mode of the EPID. After correcting for this time delay, 75% of the header angles, on average, were within ± 1° of the true gantry angle, compared to an average of only 35% without the correction. Applying a boxcar smoothing filter to the corrected gantry angles further improved the accuracy of the header-derived gantry angles to within ± 1° for almost all images (99.4%). An angle accuracy of 0.11° ± 0.04° was determined using a point-by-point comparison of the gantry angle data in the MLC DynaLog file and the potentiometer data. These simple correction methods can be easily applied to individual treatment EPID images in order to more accurately define the gantry angle.

Citing Articles

Verification of the delivered patient radiation dose for non-coplanar beam therapy.

Kutuzov I, Van Beek T, McCurdy B J Appl Clin Med Phys. 2021; 22(7):110-120.

PMID: 34021713 PMC: 8292689. DOI: 10.1002/acm2.13280.


An electronic portal image device (EPID)-based multiplatform rapid daily LINAC QA tool.

Ma Y, Wang X, Mai R, Wang T, Pei Y, Liu S J Appl Clin Med Phys. 2021; 22(1):45-58.

PMID: 33410254 PMC: 7856503. DOI: 10.1002/acm2.13055.


Evaluation of gantry angle during respiratory-gated VMAT using triggered kilovoltage x-ray image.

Miyasaka R, Saitoh H, Kawachi T, Katayose T, Cho S, Yamauchi R J Appl Clin Med Phys. 2019; 20(8):98-104.

PMID: 31355984 PMC: 6698769. DOI: 10.1002/acm2.12683.


Reducing the tracking drift of an uncontoured tumor for a portal-image-based dynamically adapted conformal radiotherapy treatment.

Teo P, Guo K, Fontaine G, Ahmed B, Alayoubi N, Kehler K Med Biol Eng Comput. 2019; 57(8):1657-1672.

PMID: 31089863 DOI: 10.1007/s11517-019-01981-4.


A novel and independent method for time-resolved gantry angle quality assurance for VMAT.

Fuangrod T, Greer P, Zwan B, Barnes M, Lehmann J J Appl Clin Med Phys. 2017; 18(5):134-142.

PMID: 28703451 PMC: 5874941. DOI: 10.1002/acm2.12129.


References
1.
Otto K . Volumetric modulated arc therapy: IMRT in a single gantry arc. Med Phys. 2008; 35(1):310-7. DOI: 10.1118/1.2818738. View

2.
Bakhtiari M, Kumaraswamy L, Bailey D, de Boer S, Malhotra H, Podgorsak M . Using an EPID for patient-specific VMAT quality assurance. Med Phys. 2011; 38(3):1366-73. DOI: 10.1118/1.3552925. View

3.
Adamson J, Wu Q . Independent verification of gantry angle for pre-treatment VMAT QA using EPID. Phys Med Biol. 2012; 57(20):6587-600. DOI: 10.1088/0031-9155/57/20/6587. View

4.
McCurdy B, Greer P . Dosimetric properties of an amorphous-silicon EPID used in continuous acquisition mode for application to dynamic and arc IMRT. Med Phys. 2009; 36(7):3028-39. DOI: 10.1118/1.3148822. View

5.
Klein E, Hanley J, Bayouth J, Yin F, Simon W, Dresser S . Task Group 142 report: quality assurance of medical accelerators. Med Phys. 2009; 36(9):4197-212. DOI: 10.1118/1.3190392. View