» Articles » PMID: 24423845

Reproducibility of the External Surface Position in Left-breast DIBH Radiotherapy with Spirometer-based Monitoring

Overview
Date 2014 Jan 16
PMID 24423845
Citations 20
Authors
Affiliations
Soon will be listed here.
Abstract

Deep inspiration breath hold (DIBH) in left-sided breast cancer radiotherapy treatments allows for a reduction in cardiac and pulmonary doses without compromising target coverage. The selection of the most appropriate technology for DIBH monitoring is a crucial issue. We evaluated the stability and reproducibility of DIBHs controlled by a spirometric device, by assessing the variability of the external surface position within a single DIBH (intra-DIBH) and between DIBHs performed in the same treatment session (intrafraction) or in different sessions (interfraction). The study included seven left-breast cancer patients treated with spirometer-based DIBH radiotherapy. Infrared optical tracking was used to record the 3D coordinates of seven to eleven passive markers placed on the patient's thoraco-abdominal surface during 29-43 DIBHs performed in six to eight treatment sessions. The obtained results showed displacements of the external surface between different sessions up to 6.3mm along a single direction, even at constant inspired volumes. The median value of the interfraction variability in the position of breast passive markers was 2.9 mm (range 1.9-4.8 mm) in the latero-lateral direction, 3.6 mm (range 2.2-4.6mm) in the antero-posterior direction, and 4.3mm (range 2.8-6.2 mm) in the cranio-caudal direction. There were no significant dose distribution variations for target and organs at risk with respect to the treatment plan, confirming the adequacy of the applied clinical margins (15 mm) to compensate for the measured setup uncertainties. This study demonstrates that spirometer-based control does not guarantee a stable and reproducible position of the external surface in left-breast DIBH radiotherapy, suggesting the need for more robust DIBH monitoring techniques when reduced margins and setup uncertainties are required for improving normal tissue sparing and decreasing cardiac and pulmonary toxicity.

Citing Articles

Prospective Randomized Trial Comparing 2 Devices for Deep Inspiration Breath Hold Management in Breast Radiation Therapy: Results of the BRAVEHeart Trial.

Byrne H, Steiner E, Booth J, Lamoury G, Morgia M, Carroll S Adv Radiat Oncol. 2024; 9(9):101572.

PMID: 39221134 PMC: 11364044. DOI: 10.1016/j.adro.2024.101572.


A method for beam's eye view breath-hold monitoring during breast volumetric modulated arc therapy.

Carr M, Gargett M, Stanton C, Zwan B, Byrne H, Booth J Phys Imaging Radiat Oncol. 2023; 25:100419.

PMID: 36875326 PMC: 9975298. DOI: 10.1016/j.phro.2023.100419.


Reproducibility and stability of spirometer-guided deep inspiration breath-hold in left-breast treatments using an optical surface monitoring system.

Cilla S, Romano C, Craus M, Viola P, Macchia G, Boccardi M J Appl Clin Med Phys. 2023; 24(6):e13922.

PMID: 36852489 PMC: 10243327. DOI: 10.1002/acm2.13922.


Feasibility and efficacy of active breathing coordinator assisted deep inspiration breath hold technique for treatment of locally advanced breast cancer.

All S, Zhao B, Montalvo S, Maxwell C, Johns C, Gu X J Appl Clin Med Phys. 2022; 24(2):e13893.

PMID: 36585853 PMC: 9924105. DOI: 10.1002/acm2.13893.


Evaluation of image-guided and surface-guided radiotherapy for breast cancer patients treated in deep inspiration breath-hold: A single institution experience.

Penninkhof J, Fremeijer K, Offereins-van Harten K, van Wanrooij C, Quint S, Kunnen B Tech Innov Patient Support Radiat Oncol. 2022; 21:51-57.

PMID: 35243045 PMC: 8861395. DOI: 10.1016/j.tipsro.2022.02.001.


References
1.
Wong V, Tung S, Ng A, Li F, Leung J . Real-time monitoring and control on deep inspiration breath-hold for lung cancer radiotherapy--combination of ABC and external marker tracking. Med Phys. 2010; 37(9):4673-83. DOI: 10.1118/1.3476463. View

2.
Moran J, Balter J, Ben-David M, Marsh R, Van Herk M, Pierce L . Short-term displacement and reproducibility of the breast and nodal targets under active breathing control. Int J Radiat Oncol Biol Phys. 2007; 68(2):541-6. PMC: 2703499. DOI: 10.1016/j.ijrobp.2007.01.028. View

3.
Clarke M, Collins R, Darby S, Davies C, Elphinstone P, Evans V . Effects of radiotherapy and of differences in the extent of surgery for early breast cancer on local recurrence and 15-year survival: an overview of the randomised trials. Lancet. 2005; 366(9503):2087-106. DOI: 10.1016/S0140-6736(05)67887-7. View

4.
Mah D, Hanley J, Rosenzweig K, Yorke E, Braban L, Ling C . Technical aspects of the deep inspiration breath-hold technique in the treatment of thoracic cancer. Int J Radiat Oncol Biol Phys. 2000; 48(4):1175-85. DOI: 10.1016/s0360-3016(00)00747-1. View

5.
Vikstrom J, Hjelstuen M, Mjaaland I, Dybvik K . Cardiac and pulmonary dose reduction for tangentially irradiated breast cancer, utilizing deep inspiration breath-hold with audio-visual guidance, without compromising target coverage. Acta Oncol. 2010; 50(1):42-50. DOI: 10.3109/0284186X.2010.512923. View