Optical Artefact Characterization and Correction in Volumetric Scintillation Dosimetry

Overview
Journal Phys Med Biol
Publisher IOP Publishing
Date 2013 Dec 11
PMID 24321820
Citations 12
Authors
Affiliations
Soon will be listed here.
Abstract

The goals of this study were (1) to characterize the optical artefacts affecting measurement accuracy in a volumetric liquid scintillator detector, and (2) to develop methods to correct for these artefacts. The optical artefacts addressed were photon scattering, refraction, camera perspective, vignetting, lens distortion, the lens point spread function, stray radiation, and noise in the camera. These artefacts were evaluated by theoretical and experimental means, and specific correction strategies were developed for each artefact. The effectiveness of the correction methods was evaluated by comparing raw and corrected images of the scintillation light from proton pencil beams against validated Monte Carlo calculations. Blurring due to the lens and refraction at the scintillator tank-air interface were found to have the largest effect on the measured light distribution, and lens aberrations and vignetting were important primarily at the image edges. Photon scatter in the scintillator was not found to be a significant source of artefacts. The correction methods effectively mitigated the artefacts, increasing the average gamma analysis pass rate from 66% to 98% for gamma criteria of 2% dose difference and 2 mm distance to agreement. We conclude that optical artefacts cause clinically meaningful errors in the measured light distribution, and we have demonstrated effective strategies for correcting these optical artefacts.

Citing Articles

Development and application of a novel scintillation gel-based 3D dosimetry system for radiotherapy.

Li H, Jin H, He L, Yan X, Zhang H, Li D J Appl Clin Med Phys. 2024; 26(3):e14615.

PMID: 39704638 PMC: 11905255. DOI: 10.1002/acm2.14615.


Imaging lung tumor motion using integrated-mode proton radiography-A phantom study towards tumor tracking in proton radiotherapy.

Fullarton R, Simard M, Volz L, Toltz A, Chung S, Schuy C Med Phys. 2024; 52(2):1146-1158.

PMID: 39530503 PMC: 11788258. DOI: 10.1002/mp.17508.


Development of a three-dimensional scintillation detector for pencil beam verification in proton therapy patient-specific quality assurance.

Frelin A, Daviau G, Bui M, Fontbonne C, Fontbonne J, Lebhertz D Med Phys. 2024; 51(12):9318-9329.

PMID: 39255360 PMC: 11656293. DOI: 10.1002/mp.17388.


Direct in-water radiation dose measurements using Cherenkov emission corrected signals from polarization imaging for a clinical radiotherapy application.

Cloutier E, Beaulieu L, Archambault L Sci Rep. 2022; 12(1):9608.

PMID: 35688843 PMC: 9187683. DOI: 10.1038/s41598-022-12672-w.


A novel proton-integrating radiography system design using a monolithic scintillator detector: experimental studies.

Darne C, Robertson D, Alsanea F, Collins-Fekete C, Beddar S Nucl Instrum Methods Phys Res A. 2022; 1027.

PMID: 35221402 PMC: 8872121. DOI: 10.1016/j.nima.2021.166077.


References
1.
Archambault L, Poenisch F, Sahoo N, Robertson D, Lee A, Gillin M . Verification of proton range, position, and intensity in IMPT with a 3D liquid scintillator detector system. Med Phys. 2012; 39(3):1239-46. PMC: 3292596. DOI: 10.1118/1.3681948. View

2.
Biggs D, Andrews M . Acceleration of iterative image restoration algorithms. Appl Opt. 1997; 36(8):1766-75. DOI: 10.1364/ao.36.001766. View

3.
Low D, Harms W, Mutic S, Purdy J . A technique for the quantitative evaluation of dose distributions. Med Phys. 1998; 25(5):656-61. DOI: 10.1118/1.598248. View

4.
Robertson D, Mirkovic D, Sahoo N, Beddar S . Quenching correction for volumetric scintillation dosimetry of proton beams. Phys Med Biol. 2012; 58(2):261-73. PMC: 3849813. DOI: 10.1088/0031-9155/58/2/261. View

5.
Kirov A, Piao J, Mathur N, Miller T, Devic S, Trichter S . The three-dimensional scintillation dosimetry method: test for a 106Ru eye plaque applicator. Phys Med Biol. 2005; 50(13):3063-81. DOI: 10.1088/0031-9155/50/13/007. View