Experimental Validation of the TOPAS Monte Carlo System for Passive Scattering Proton Therapy

Overview
Journal Med Phys
Specialty Biophysics
Date 2013 Dec 11
PMID 24320505
Citations 37
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: TOPAS (TOol for PArticle Simulation) is a particle simulation code recently developed with the specific aim of making Monte Carlo simulations user-friendly for research and clinical physicists in the particle therapy community. The authors present a thorough and extensive experimental validation of Monte Carlo simulations performed with TOPAS in a variety of setups relevant for proton therapy applications. The set of validation measurements performed in this work represents an overall end-to-end testing strategy recommended for all clinical centers planning to rely on TOPAS for quality assurance or patient dose calculation and, more generally, for all the institutions using passive-scattering proton therapy systems.

Methods: The authors systematically compared TOPAS simulations with measurements that are performed routinely within the quality assurance (QA) program in our institution as well as experiments specifically designed for this validation study. First, the authors compared TOPAS simulations with measurements of depth-dose curves for spread-out Bragg peak (SOBP) fields. Second, absolute dosimetry simulations were benchmarked against measured machine output factors (OFs). Third, the authors simulated and measured 2D dose profiles and analyzed the differences in terms of field flatness and symmetry and usable field size. Fourth, the authors designed a simple experiment using a half-beam shifter to assess the effects of multiple Coulomb scattering, beam divergence, and inverse square attenuation on lateral and longitudinal dose profiles measured and simulated in a water phantom. Fifth, TOPAS' capabilities to simulate time dependent beam delivery was benchmarked against dose rate functions (i.e., dose per unit time vs time) measured at different depths inside an SOBP field. Sixth, simulations of the charge deposited by protons fully stopping in two different types of multilayer Faraday cups (MLFCs) were compared with measurements to benchmark the nuclear interaction models used in the simulations.

Results: SOBPs' range and modulation width were reproduced, on average, with an accuracy of +1, -2 and ±3 mm, respectively. OF simulations reproduced measured data within ±3%. Simulated 2D dose-profiles show field flatness and average field radius within ±3% of measured profiles. The field symmetry resulted, on average in ±3% agreement with commissioned profiles. TOPAS accuracy in reproducing measured dose profiles downstream the half beam shifter is better than 2%. Dose rate function simulation reproduced the measurements within ∼2% showing that the four-dimensional modeling of the passively modulation system was implement correctly and millimeter accuracy can be achieved in reproducing measured data. For MLFCs simulations, 2% agreement was found between TOPAS and both sets of experimental measurements. The overall results show that TOPAS simulations are within the clinical accepted tolerances for all QA measurements performed at our institution.

Conclusions: Our Monte Carlo simulations reproduced accurately the experimental data acquired through all the measurements performed in this study. Thus, TOPAS can reliably be applied to quality assurance for proton therapy and also as an input for commissioning of commercial treatment planning systems. This work also provides the basis for routine clinical dose calculations in patients for all passive scattering proton therapy centers using TOPAS.

Citing Articles

Proton Beam Range and Charge Verification Using Multilayer Faraday Collector.

Yeap P, Lew K, Koh W, Chua C, Wibawa A, Master Z Technol Cancer Res Treat. 2024; 23:15330338241262610.

PMID: 39051529 PMC: 11320672. DOI: 10.1177/15330338241262610.


High-Density Glass Scintillators for Proton Radiography-Relative Luminosity, Proton Response, and Spatial Resolution.

Stolen E, Fullarton R, Hein R, Conner R, Jacobsohn L, Collins-Fekete C Sensors (Basel). 2024; 24(7).

PMID: 38610351 PMC: 11014246. DOI: 10.3390/s24072137.


Experimental validation of proton physics models of Geant4 for calculating stopping power ratio.

Liu R, Zhao X, Medrano M J Radiol Prot. 2022; 42(2).

PMID: 35705062 PMC: 9462414. DOI: 10.1088/1361-6498/ac7918.


Quantification of biological range uncertainties in patients treated at the Krakow proton therapy centre.

Garbacz M, Gajewski J, Durante M, Kisielewicz K, Krah N, Kopec R Radiat Oncol. 2022; 17(1):50.

PMID: 35264184 PMC: 8905899. DOI: 10.1186/s13014-022-02022-5.


Dosimetry and radioprotection evaluations of very high energy electron beams.

Masilela T, Delorme R, Prezado Y Sci Rep. 2021; 11(1):20184.

PMID: 34642417 PMC: 8511248. DOI: 10.1038/s41598-021-99645-7.


References
1.
Kooy H, Rosenthal S, Engelsman M, Mazal A, Slopsema R, Paganetti H . The prediction of output factors for spread-out proton Bragg peak fields in clinical practice. Phys Med Biol. 2005; 50(24):5847-56. DOI: 10.1088/0031-9155/50/24/006. View

2.
Arjomandy B, Sahoo N, Ding X, Gillin M . Use of a two-dimensional ionization chamber array for proton therapy beam quality assurance. Med Phys. 2008; 35(9):3889-94. DOI: 10.1118/1.2963990. View

3.
Ramos-Mendez J, Perl J, Faddegon B, Schumann J, Paganetti H . Geometrical splitting technique to improve the computational efficiency in Monte Carlo calculations for proton therapy. Med Phys. 2013; 40(4):041718. PMC: 3618101. DOI: 10.1118/1.4795343. View

4.
Schumann J, Paganetti H, Shin J, Faddegon B, Perl J . Efficient voxel navigation for proton therapy dose calculation in TOPAS and Geant4. Phys Med Biol. 2012; 57(11):3281-93. PMC: 3367506. DOI: 10.1088/0031-9155/57/11/3281. View

5.
Perl J, Shin J, Schumann J, Faddegon B, Paganetti H . TOPAS: an innovative proton Monte Carlo platform for research and clinical applications. Med Phys. 2012; 39(11):6818-37. PMC: 3493036. DOI: 10.1118/1.4758060. View