» Articles » PMID: 24165395

Development of a Transportable Neutron Activation Analysis System to Quantify Manganese in Bone in Vivo: Feasibility and Methodology

Overview
Journal Physiol Meas
Date 2013 Oct 30
PMID 24165395
Citations 19
Authors
Affiliations
Soon will be listed here.
Abstract

This study was conducted to investigate the methodology and feasibility of developing a transportable neutron activation analysis (NAA) system to quantify manganese (Mn) in bone using a portable deuterium-deuterium (DD) neutron generator as the neutron source. Since a DD neutron generator was not available in our laboratory, a deuterium-tritium (DT) neutron generator was used to obtain experimental data and validate the results from Monte Carlo (MC) simulations. After validation, MC simulations using a DD generator as the neutron source were then conducted. Different types of moderators and reflectors were simulated, and the optimal thicknesses for the moderator and reflector were determined. To estimate the detection limit (DL) of the system, and to observe the interference of the magnesium (Mg) γ line at 844 keV to the Mn γ line at 847 keV, three hand phantoms with Mn concentrations of 30 parts per million (ppm), 150 ppm, and 500 ppm were made and irradiated by the DT generator system. The Mn signals in these phantoms were then measured using a 50% high-efficiency high-purity germanium (HPGe) detector. The DL was calculated to be about 4.4 ppm for the chosen irradiation, decay, and measurement time. This was calculated to be equivalent to a DL of about 3.3 ppm for the DD generator system. To achieve this DL with one 50% high-efficiency HPGe detector, the dose to the hand was simulated to be about 37 mSv, with the total body equivalent dose being about 23µSv. In conclusion, it is feasible to develop a transportable NAA system to quantify Mn in bone in vivo with an acceptable radiation exposure to the subject.

Citing Articles

Sub-micrometer scale synchrotron x-ray fluorescence measurements of trace elements in teeth compared with laser ablation inductively coupled plasma mass spectrometry.

Specht A, Zhang X, Antipova O, Sayam A, Nguyen V, Hoover C J Expo Sci Environ Epidemiol. 2025; .

PMID: 39881199 DOI: 10.1038/s41370-025-00754-6.


The association of bone and blood manganese with motor function in Chinese workers.

Rolle-McFarland D, Liu Y, Mostafaei F, Zauber S, Zhou Y, Li Y Neurotoxicology. 2021; 88:224-230.

PMID: 34896555 PMC: 8748420. DOI: 10.1016/j.neuro.2021.12.005.


Bone manganese is a sensitive biomarker of ongoing elevated manganese exposure, but does not accumulate across the lifespan.

Conley T, Richardson C, Pacheco J, Dave N, Jursa T, Guazzetti S Environ Res. 2021; 204(Pt D):112355.

PMID: 34774504 PMC: 10413361. DOI: 10.1016/j.envres.2021.112355.


Molecular and physical technologies for monitoring fluid and electrolyte imbalance: A focus on cancer population.

Bennet D, Khorsandian Y, Pelusi J, Mirabella A, Pirrotte P, Zenhausern F Clin Transl Med. 2021; 11(6):e461.

PMID: 34185420 PMC: 8214861. DOI: 10.1002/ctm2.461.


Biodegradation, Antibacterial Performance, and Cytocompatibility of a Novel ZK30-Cu-Mn Biomedical Alloy Produced by Selective Laser Melting.

Xie B, Zhao M, Xu R, Zhao Y, Yin D, Gao C Int J Bioprint. 2021; 7(1):300.

PMID: 33585710 PMC: 7875054. DOI: 10.18063/ijb.v7i1.300.


References
1.
Bowman A, Kwakye G, Herrero Hernandez E, Aschner M . Role of manganese in neurodegenerative diseases. J Trace Elem Med Biol. 2011; 25(4):191-203. PMC: 3230726. DOI: 10.1016/j.jtemb.2011.08.144. View

2.
Wyatt R, Ryde S, Morgan W, McNeil E, Hainsworth I, Williams A . The development of a technique to measure bone aluminium content using neutron activation analysis. Physiol Meas. 1993; 14(3):327-35. DOI: 10.1088/0967-3334/14/3/011. View

3.
Smith D, Gwiazda R, Bowler R, Roels H, Park R, Taicher C . Biomarkers of Mn exposure in humans. Am J Ind Med. 2007; 50(11):801-11. DOI: 10.1002/ajim.20506. View

4.
Qin W, Fu X, Jiang Y, Long L, Li X, Chen H . [Variations of brain magnetic resonance imaging among manganese-exposed workers]. Zhonghua Yu Fang Yi Xue Za Zhi. 2010; 43(9):793-7. View

5.
Seaborn C, Nielsen F . Dietary silicon and arginine affect mineral element composition of rat femur and vertebra. Biol Trace Elem Res. 2002; 89(3):239-50. DOI: 10.1385/bter:89:3:239. View