Lumped Element Electrical Model Based on Three Resistors for Electrical Impedance in Radiofrequency Cardiac Ablation: Estimations from Analytical Calculations and Clinical Data

Overview
Publisher Bentham Open
Date 2013 Aug 21
PMID 23961299
Citations 5
Authors
Affiliations
Soon will be listed here.
Abstract

The electrical impedance measured during radiofrequency cardiac ablation (RFCA) is widely used in clinical studies to predict the heating evolution and hence the success of the procedure. We hypothesized that a model based on three resistors in series can mimic the total electrical impedance measured during RFCA. The three resistors or impedances are given by: impedance associated with the tissue around the active electrode (myocardium and circulating blood) (Z-A), that associated with the tissue around the dispersive electrode (Z-DE) and that associated with the rest of the body (Z-B). Our objective was to quantify the values associated with these three impedance types by an analytical method, after which the values obtained would be compared to those estimated from clinical data from previous studies. The results suggest that an RFCA using a 7 Fr 4-mm electrode would give a Z-A of around 75 ohms, a Z-DE around 20 ohms, and Z-B would be 15±10 ohms (for body surface area variations between 1.5 and 2.5 m^2). Finally, adaptations of the proposed model were used to explain the results of previous clinical studies using a different electrode arrangement, such as in bipolar ablation of the ventricular septum.

Citing Articles

Evaluation of lesion characteristics and baseline impedance on high-power short-duration radiofrequency catheter ablation using computer simulation.

Sun Y, Zhu X, Nakamura K, Wang S Heart Vessels. 2023; 38(12):1459-1467.

PMID: 37650926 DOI: 10.1007/s00380-023-02300-6.


Limitations of Baseline Impedance, Impedance Drop and Current for Radiofrequency Catheter Ablation Monitoring: Insights from In silico Modeling.

Irastorza R, Maher T, Barkagan M, Liubasuskas R, Perez J, Berjano E J Cardiovasc Dev Dis. 2022; 9(10).

PMID: 36286288 PMC: 9604830. DOI: 10.3390/jcdd9100336.


Effect of Baseline Impedance in Radiofrequency Delivery on Lesion Characteristics and the Relationship Between Impedance and Steam Pops.

Qu L, Guo M, Sun M, Wang R, Zhang N, Li X Front Cardiovasc Med. 2022; 9:872961.

PMID: 35571193 PMC: 9094712. DOI: 10.3389/fcvm.2022.872961.


Impact of baseline impedance of pulmonary vein antrum on success of catheter ablation for paroxysmal atrial fibrillation guided by ablation index.

Sun Y, Xiao X, Yin X, Gao L, Yu X, Zhang R BMC Cardiovasc Disord. 2022; 22(1):179.

PMID: 35439961 PMC: 9016960. DOI: 10.1186/s12872-022-02530-y.


Should fluid dynamics be included in computer models of RF cardiac ablation by irrigated-tip electrodes?.

Gonzalez-Suarez A, Perez J, Berjano E Biomed Eng Online. 2018; 17(1):43.

PMID: 29678186 PMC: 5910590. DOI: 10.1186/s12938-018-0475-7.

References
1.
Ragheb T, Riegle S, Geddes L, Amin V . The impedance of a spherical monopolar electrode. Ann Biomed Eng. 1992; 20(6):617-27. DOI: 10.1007/BF02368609. View

2.
Panescu D, Whayne J, Fleischman S, Mirotznik M, Swanson D, Webster J . Three-dimensional finite element analysis of current density and temperature distributions during radio-frequency ablation. IEEE Trans Biomed Eng. 1995; 42(9):879-90. DOI: 10.1109/10.412649. View

3.
Koruth J, Dukkipati S, Miller M, Neuzil P, dAvila A, Reddy V . Bipolar irrigated radiofrequency ablation: a therapeutic option for refractory intramural atrial and ventricular tachycardia circuits. Heart Rhythm. 2012; 9(12):1932-41. DOI: 10.1016/j.hrthm.2012.08.001. View

4.
Wang D, Hulse J, Walsh E, Saul J . Factors influencing impedance during radiofrequency ablation in humans. Chin Med J (Engl). 1995; 108(6):450-5. View

5.
Wittkampf F, Nakagawa H . RF catheter ablation: Lessons on lesions. Pacing Clin Electrophysiol. 2006; 29(11):1285-97. DOI: 10.1111/j.1540-8159.2006.00533.x. View