Body Composition Assessment in Overweight Women: Validation of Air Displacement Plethysmography

Overview
Date 2013 Jul 17
PMID 23855413
Citations 13
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: The purpose of this study was to evaluate the validity and reliability of air displacement plethysmography (ADP) compared to a dual energy x-ray absorptiometry (DXA) criterion for body composition measurement in overweight and obese women (BMI ≥ 25·0 kg m(2) ).

Subjects/methods: Twenty-four overweight and obese women (Mean ± SD; Age: 36·6 ± 12·0 years; Height: 166·4 ± 5·8 cm; Weight: 86·5 ± 14·2 kg; Body Fat: 38·5 ± 3·7%; BMI: 31·3 ± 5·5 kg m(2) ) were tested after an 8-h fast. Fat mass (FM), fat-free mass (FFM) and percent body fat (%BF) were measured by ADP and compared to values determined by the DXA criterion. FFM from DXA was calculated as lean mass plus bone mineral content. A paired samples t-test was used to test for significant differences in the body composition variables between methods. A one-way ANOVA along with intraclass correlation coefficient (ICC), SEM,%SEM and MD was used to represent reliability.

Results: Validity data comparing ADP and DXA demonstrated no significant difference in FM (ADP-DXA FM = 0·99 kg; P = 0·113), FFM (0·98 kg; P = 0·115) and %BF (1·56%; P = 0·540). Reliability data for ADP between the first and second trials showed no significant difference in FM (P = 0·168; ICC = 0·994; SEM = 0·668), FFM (P = 0·058; ICC = 0·973; SEM = 0·892) or %BF (P = 0·121; ICC = 0·971; SEM = 0·813).

Conclusions: For overweight and obese women, ADP was found to be a valid measure of FM, FFM and %BF when compared with DXA. The reliability of ADP was supported for all body composition variables.

Citing Articles

Associations between 47 anthropometric markers derived from a body scanner and relative fat-free mass in a population-based study.

Dietzmann M, Radke D, Markus M, Wiese M, Volzke H, Felix S BMC Public Health. 2024; 24(1):1079.

PMID: 38637778 PMC: 11025281. DOI: 10.1186/s12889-024-18611-w.


The Preventiometer - reliability of a cardiovascular multi-device measurement platform and its measurement agreement with a cohort study.

Junge M, Kruger M, Wahner-Roedler D, Bauer B, Dorr M, Bahls M BMC Med Res Methodol. 2023; 23(1):103.

PMID: 37095457 PMC: 10127382. DOI: 10.1186/s12874-023-01911-x.


Modeling Risk Factors for Sleep- and Adiposity-Related Cardiometabolic Disease: Protocol for the Short Sleep Undermines Cardiometabolic Health (SLUMBRx) Observational Study.

Knowlden A, Higginbotham J, Grandner M, Allegrante J JMIR Res Protoc. 2021; 10(3):e27139.

PMID: 33687340 PMC: 7988396. DOI: 10.2196/27139.


Validity of Body-Composition Methods across Racial and Ethnic Populations.

Blue M, Tinsley G, Ryan E, Smith-Ryan A Adv Nutr. 2021; 12(5):1854-1862.

PMID: 33684215 PMC: 8528114. DOI: 10.1093/advances/nmab016.


Dietary Patterns in New Zealand Women: Evaluating Differences in Body Composition and Metabolic Biomarkers.

Jayasinghe S, Breier B, McNaughton S, Russell A, Della Gatta P, Mason S Nutrients. 2019; 11(7).

PMID: 31323812 PMC: 6682986. DOI: 10.3390/nu11071643.


References
1.
Brownbill R, Ilich J . Measuring body composition in overweight individuals by dual energy x-ray absorptiometry. BMC Med Imaging. 2005; 5(1):1. PMC: 1079847. DOI: 10.1186/1471-2342-5-1. View

2.
Hopkins W . Measures of reliability in sports medicine and science. Sports Med. 2000; 30(1):1-15. DOI: 10.2165/00007256-200030010-00001. View

3.
Bertoli S, Battezzati A, Testolin G, Bedogni G . Evaluation of air-displacement plethysmography and bioelectrical impedance analysis vs dual-energy X-ray absorptiometry for the assessment of fat-free mass in elderly subjects. Eur J Clin Nutr. 2007; 62(11):1282-6. DOI: 10.1038/sj.ejcn.1602847. View

4.
Minderico C, Silva A, Teixeira P, Sardinha L, Hull H, Fields D . Validity of air-displacement plethysmography in the assessment of body composition changes in a 16-month weight loss program. Nutr Metab (Lond). 2006; 3:32. PMC: 1560140. DOI: 10.1186/1743-7075-3-32. View

5.
Edwards H, Randall Simpson J, Buchholz A . Air displacement plethysmography for fat-mass measurement in healthy young women. Can J Diet Pract Res. 2011; 72(2):85-7. DOI: 10.3148/72.2.2011.85. View