Cell Counting of Body Fluids: Comparison Between Three Automated Haematology Analysers and the Manual Microscope Method

Overview
Specialty Hematology
Date 2013 May 8
PMID 23647736
Citations 7
Authors
Affiliations
Soon will be listed here.
Abstract

Introduction: Haematological analysis of body fluids (BF) specimens can provide clinicians with valuable diagnostic information because it can indicate one of several serious medical conditions. Although up to now the microscopic counting and the differentiation of WBC in a BF smear have been used as a reference. The introduction of semiautomated and automated methods of analysis has reduced interoperator variability and improved turnaround time and precision. The aim of our study was to evaluate the accuracy and the correlation between the three methods and with the reference method.

Methods: We examined 110 body fluid samples. Total counting of each sample has been conducted with all systems: Pentra DX120, ADVIA 2120 and XE-2100 and the manual method.

Results: We found statistically significant correlation between the data obtained in the ascitic and pleuric liquid but not in the cerebrospinal fluid.

Conclusion: The introduction of automated method for BF analysis is more and more useful in the routine job of a laboratory analysis. It is therefore very important to evaluate the performance of the different automated haematology technologies, because there is a lack of literature in this field. The comparison between the Pentra DX 120, the other technologies and the manual counting showed instrumental overlapping capabilities.

Citing Articles

Canine Cerebrospinal Fluid Analysis Using Two New Automated Techniques: The Sysmex XN-V Body Fluid Mode and an Artificial-Intelligence-Based Algorithm.

Lapsina S, Riond B, Hofmann-Lehmann R, Stirn M Animals (Basel). 2024; 14(11).

PMID: 38891702 PMC: 11171029. DOI: 10.3390/ani14111655.


Comparison of Sysmex XN-V body fluid mode and deep-learning-based quantification with manual techniques for total nucleated cell count and differential count for equine bronchoalveolar lavage samples.

Lapsina S, Riond B, Hofmann-Lehmann R, Stirn M BMC Vet Res. 2024; 20(1):48.

PMID: 38317167 PMC: 10840287. DOI: 10.1186/s12917-024-03884-5.


Automated cell count in body fluids: a review.

Alcaide Martin M, Altimira Queral L, Sahuquillo Frias L, Valina Amado L, Merino A, de Guadiana-Romualdo L Adv Lab Med. 2023; 2(2):149-177.

PMID: 37363326 PMC: 10197423. DOI: 10.1515/almed-2021-0011.


A New Graphic Type Differentiation of Cell Account Determination for Distinguishing Acute Periprosthetic Joint Infection from Hemarthrosis.

Fink B, Hoyka M, Weissbarth E, Schuster P, Berger I Antibiotics (Basel). 2022; 11(10).

PMID: 36289943 PMC: 9598683. DOI: 10.3390/antibiotics11101284.


Performance Evaluation of Body Fluid Cellular Analysis Using the Beckman Coulter UniCel DxH 800, Sysmex XN-350, and UF-5000 Automated Cellular Analyzers.

Cho J, Oh J, Lee S, Lee Y, Song J, Kim J Ann Lab Med. 2019; 40(2):122-130.

PMID: 31650728 PMC: 6822009. DOI: 10.3343/alm.2020.40.2.122.