Prediction of Human Observer Performance in a 2-alternative Forced Choice Low-contrast Detection Task Using Channelized Hotelling Observer: Impact of Radiation Dose and Reconstruction Algorithms

Overview
Journal Med Phys
Specialty Biophysics
Date 2013 Apr 6
PMID 23556902
Citations 69
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: Efficient optimization of CT protocols demands a quantitative approach to predicting human observer performance on specific tasks at various scan and reconstruction settings. The goal of this work was to investigate how well a channelized Hotelling observer (CHO) can predict human observer performance on 2-alternative forced choice (2AFC) lesion-detection tasks at various dose levels and two different reconstruction algorithms: a filtered-backprojection (FBP) and an iterative reconstruction (IR) method.

Methods: A 35 × 26 cm(2) torso-shaped phantom filled with water was used to simulate an average-sized patient. Three rods with different diameters (small: 3 mm; medium: 5 mm; large: 9 mm) were placed in the center region of the phantom to simulate small, medium, and large lesions. The contrast relative to background was -15 HU at 120 kV. The phantom was scanned 100 times using automatic exposure control each at 60, 120, 240, 360, and 480 quality reference mAs on a 128-slice scanner. After removing the three rods, the water phantom was again scanned 100 times to provide signal-absent background images at the exact same locations. By extracting regions of interest around the three rods and on the signal-absent images, the authors generated 21 2AFC studies. Each 2AFC study had 100 trials, with each trial consisting of a signal-present image and a signal-absent image side-by-side in randomized order. In total, 2100 trials were presented to both the model and human observers. Four medical physicists acted as human observers. For the model observer, the authors used a CHO with Gabor channels, which involves six channel passbands, five orientations, and two phases, leading to a total of 60 channels. The performance predicted by the CHO was compared with that obtained by four medical physicists at each 2AFC study.

Results: The human and model observers were highly correlated at each dose level for each lesion size for both FBP and IR. The Pearson's product-moment correlation coefficients were 0.986 [95% confidence interval (CI): 0.958-0.996] for FBP and 0.985 (95% CI: 0.863-0.998) for IR. Bland-Altman plots showed excellent agreement for all dose levels and lesions sizes with a mean absolute difference of 1.0% ± 1.1% for FBP and 2.1% ± 3.3% for IR.

Conclusions: Human observer performance on a 2AFC lesion detection task in CT with a uniform background can be accurately predicted by a CHO model observer at different radiation dose levels and for both FBP and IR methods.

Citing Articles

State-of-the-Art Deep Learning CT Reconstruction Algorithms in Abdominal Imaging.

Mileto A, Yu L, Revels J, Kamel S, Shehata M, Ibarra-Rovira J Radiographics. 2024; 44(12):e240095.

PMID: 39612283 PMC: 11618294. DOI: 10.1148/rg.240095.


Low-contrast detectability of photon-counting-detector CT at different scan modes and image types in comparison with energy-integrating-detector CT.

Fan M, Zhou Z, Wellinghoff J, McCollough C, Yu L J Med Imaging (Bellingham). 2024; 11(Suppl 1):S12803.

PMID: 38799271 PMC: 11116128. DOI: 10.1117/1.JMI.11.S1.S12803.


Experimental measurement of local noise power spectrum (NPS) in photon counting detector-CT (PCD-CT) using a single data acquisition.

Zhang C, Li K, Zhang R, Chen G Med Phys. 2024; 51(6):4081-4094.

PMID: 38703355 PMC: 11147724. DOI: 10.1002/mp.17110.


Automated Web-based Software for CT Quality Control Testing of Low-contrast Detectability using Model Observers.

Zhou Z, Wellinghoff J, Fan M, Hsieh S, Holmes 3rd D, McCollough C Proc SPIE Int Soc Opt Eng. 2024; 12925.

PMID: 38606000 PMC: 11008424. DOI: 10.1117/12.3008777.


3D printed phantom with 12 000 submillimeter lesions to improve efficiency in CT detectability assessment.

Shunhavanich P, Mei K, Shapira N, Stayman J, McCollough C, Gang G Med Phys. 2024; 51(5):3265-3274.

PMID: 38588491 PMC: 11076156. DOI: 10.1002/mp.17064.


References
1.
Fetterly K, Blume H, Flynn M, Samei E . Introduction to grayscale calibration and related aspects of medical imaging grade liquid crystal displays. J Digit Imaging. 2007; 21(2):193-207. PMC: 3043865. DOI: 10.1007/s10278-007-9022-y. View

2.
Gang G, Tward D, Lee J, Siewerdsen J . Anatomical background and generalized detectability in tomosynthesis and cone-beam CT. Med Phys. 2010; 37(5):1948-65. PMC: 2862054. DOI: 10.1118/1.3352586. View

3.
McCollough C, Chen G, Kalender W, Leng S, Samei E, Taguchi K . Achieving routine submillisievert CT scanning: report from the summit on management of radiation dose in CT. Radiology. 2012; 264(2):567-80. PMC: 3401354. DOI: 10.1148/radiol.12112265. View

4.
Eckstein M, Whiting J . Lesion detection in structured noise. Acad Radiol. 1995; 2(3):249-53. DOI: 10.1016/s1076-6332(05)80174-6. View

5.
Richard S, Siewerdsen J . Comparison of model and human observer performance for detection and discrimination tasks using dual-energy x-ray images. Med Phys. 2008; 35(11):5043-53. PMC: 2673596. DOI: 10.1118/1.2988161. View