Acetabular Liner with Focal Constraint to Prevent Dislocation After THA

Overview
Publisher Wolters Kluwer
Specialty Orthopedics
Date 2013 Feb 21
PMID 23423623
Citations 8
Authors
Affiliations
Soon will be listed here.
Abstract

Background: Dislocation continues to commonly cause failure after primary and revision total hip arthroplasty (THA). Fully constrained liners intended to prevent dislocation are nonetheless associated with a substantial incidence of failure by redislocation, mechanical failure, aseptic loosening, or a combination. Constrained liners with cutouts of the elevated rims can theoretically increase range of movement and therefore decrease the risk dislocation, but it is unclear if they do so in practice and whether they are associated with early wear or loosening.

Questions/purposes: We therefore determined (1) occurrence or recurrence of dislocation and (2) rate of complications associated with constrained implants with cutouts; and (3) assessed for early cup loosening.

Methods: We retrospectively reviewed the records of 81 patients at high risk for dislocation who had 82 constrained liners inserted for primary (n = 10) or revision (n = 72) THA between 2008 and 2010. From the records we extracted demographic and implant data and instances of recurrent dislocation, implant failure, osteolysis, loosening, or construct failure. The minimum followup was 24 months (mean, 34 months; range, 24-49 months).

Results: Three liners failed as a result of further dislocation (3%). Three deep infections occurred. One patient had progressive loosening at the shell-bone interface.

Conclusions: Our observations suggest this liner is associated with a relatively low risk of dislocation in patients at high risk for dislocation and those with recurrent dislocation.

Citing Articles

Evaluating the risk of return to the operating room across all elective orthopaedic procedures.

Kiritsis N, Harris M, Reiter C, Ernst B, Satalich J, Wyatt P J Orthop Surg Res. 2024; 19(1):328.

PMID: 38825677 PMC: 11145881. DOI: 10.1186/s13018-024-04814-9.


Removing Cemented Dual Mobility Liners From Acetabular Components: A Technical Tip for a Challenging Surgical Problem.

Wilson J, Mika A, Martin J Arthroplast Today. 2023; 23:101211.

PMID: 37781023 PMC: 10541465. DOI: 10.1016/j.artd.2023.101211.


Outcomes of total hip arthroplasty with a standard prosthesis for the affected hip in patients with poliomyelitis sequelae: a mid-term retrospective follow-up study.

Shi L, Che G, Huang Y, Yi M, Kang P J Orthop Surg Res. 2023; 18(1):195.

PMID: 36915109 PMC: 10009972. DOI: 10.1186/s13018-023-03697-6.


Dual mobility acetabular construct with freedom constrained liner for treatment of recurrent dislocations after total hip arthroplasty: A case report and literature review.

Onochi Y, Fukui K, Kaneuji A, Ichiseki T, Wang X, Kawahara N Int J Surg Case Rep. 2020; 75:94-99.

PMID: 32932039 PMC: 7493044. DOI: 10.1016/j.ijscr.2020.09.012.


Comparison of Constrained Acetabular Components and Dual Mobility Cups in Revision Total Hip Arthroplasty: A Literature Review.

van Eecke E, Vanbiervliet J, Dauwe J, Mulier M Hip Pelvis. 2020; 32(2):59-69.

PMID: 32566536 PMC: 7295616. DOI: 10.5371/hp.2020.32.2.59.


References
1.
Shah S, Kaye R, Kelly N, Su E, Padgett D, Wright T . Retrieval analysis of failed constrained acetabular liners. J Arthroplasty. 2009; 24(6 Suppl):54-7. DOI: 10.1016/j.arth.2009.05.002. View

2.
Lautridou C, Lebel B, Burdin G, Vielpeau C . [Survival of the cementless Bousquet dual mobility cup: Minimum 15-year follow-up of 437 total hip arthroplasties]. Rev Chir Orthop Reparatrice Appar Mot. 2008; 94(8):731-9. DOI: 10.1016/j.rco.2008.06.001. View

3.
Havelin L, Robertsson O, Fenstad A, Overgaard S, Garellick G, Furnes O . A Scandinavian experience of register collaboration: the Nordic Arthroplasty Register Association (NARA). J Bone Joint Surg Am. 2012; 93 Suppl 3:13-9. DOI: 10.2106/JBJS.K.00951. View

4.
Shapiro G, Weiland D, Markel D, Padgett D, Sculco T, Pellicci P . The use of a constrained acetabular component for recurrent dislocation. J Arthroplasty. 2003; 18(3):250-8. DOI: 10.1054/arth.2003.50090. View

5.
Noble P, Durrani S, Usrey M, Mathis K, Bardakos N . Constrained cups appear incapable of meeting the demands of revision THA. Clin Orthop Relat Res. 2011; 470(7):1907-16. PMC: 3369098. DOI: 10.1007/s11999-011-2212-4. View