» Articles » PMID: 23366287

Performance Evaluation of a Lower Limb Exoskeleton for Stair Ascent and Descent with Paraplegia

Overview
Date 2013 Feb 1
PMID 23366287
Citations 11
Authors
Affiliations
Soon will be listed here.
Abstract

This paper describes the application of a powered lower limb exoskeleton to aid paraplegic individuals in stair ascent and descent. A brief description of the exoskeleton hardware is provided along with an explanation of the control methodology implemented to allow stair ascent and descent. Tests were performed with a paraplegic individual (T10 complete injury level) and data is presented from multiple trials, including the hip and knee joint torque and power required to perform this functionality. Joint torque and power requirements are summarized, including peak hip and knee joint torque requirements of 0.75 Nm/kg and 0.87 Nm/kg, respectively, and peak hip and knee joint power requirements of approximately 0.65 W/kg and 0.85 W/kg, respectively.

Citing Articles

The Effects of Lower Limb Orthoses on Health Aspects of the Spinal Cord Injury Patients: A Systematic Review Using International Classification of Functioning, Disability, and Health (ICF) as a Reference Framework.

Fallahzadeh Abarghuei A, Karimi M Med J Islam Repub Iran. 2023; 36:153.

PMID: 36654846 PMC: 9832937. DOI: 10.47176/mjiri.36.153.


Lower Limb Exoskeleton Sensors: State-of-the-Art.

Netukova S, Bejtic M, Mala C, Horakova L, Kutilek P, Kauler J Sensors (Basel). 2022; 22(23).

PMID: 36501804 PMC: 9738474. DOI: 10.3390/s22239091.


Converging Robotic Technologies in Targeted Neural Rehabilitation: A Review of Emerging Solutions and Challenges.

Nizamis K, Athanasiou A, Almpani S, Dimitrousis C, Astaras A Sensors (Basel). 2021; 21(6).

PMID: 33809721 PMC: 8002299. DOI: 10.3390/s21062084.


A Novel User Control for Lower Extremity Rehabilitation Exoskeletons.

Karunakaran K, Abbruzzese K, Androwis G, Foulds R Front Robot AI. 2021; 7:108.

PMID: 33501275 PMC: 7805763. DOI: 10.3389/frobt.2020.00108.


Lower limb joint biomechanics-based identification of gait transitions in between level walking and stair ambulation.

Grimmer M, Zeiss J, Weigand F, Zhao G, Lamm S, Steil M PLoS One. 2020; 15(9):e0239148.

PMID: 32936793 PMC: 7494088. DOI: 10.1371/journal.pone.0239148.


References
1.
Kobetic R, To C, Schnellenberger J, Audu M, Bulea T, Gaudio R . Development of hybrid orthosis for standing, walking, and stair climbing after spinal cord injury. J Rehabil Res Dev. 2009; 46(3):447-62. View

2.
Ohta Y, Yano H, Suzuki R, Yoshida M, Kawashima N, Nakazawa K . A two-degree-of-freedom motor-powered gait orthosis for spinal cord injury patients. Proc Inst Mech Eng H. 2007; 221(6):629-39. DOI: 10.1243/09544119JEIM55. View

3.
Neuhaus P, Noorden J, Craig T, Torres T, Kirschbaum J, Pratt J . Design and evaluation of Mina: a robotic orthosis for paraplegics. IEEE Int Conf Rehabil Robot. 2012; 2011:5975468. DOI: 10.1109/ICORR.2011.5975468. View

4.
Durfee W, Rivard A . Design and simulation of a pneumatic, stored-energy, hybrid orthosis for gait restoration. J Biomech Eng. 2006; 127(6):1014-9. DOI: 10.1115/1.2050652. View

5.
Quintero H, Farris R, Goldfarb M . Control and implementation of a powered lower limb orthosis to aid walking in paraplegic individuals. IEEE Int Conf Rehabil Robot. 2012; 2011:5975481. PMC: 3402219. DOI: 10.1109/ICORR.2011.5975481. View