Increased Blood Glycohemoglobin A1c Levels Lead to Overestimation of Arterial Oxygen Saturation by Pulse Oximetry in Patients with Type 2 Diabetes

Overview
Publisher Biomed Central
Date 2012 Sep 19
PMID 22985301
Citations 15
Authors
Affiliations
Soon will be listed here.
Abstract

Background: Non-enzymatic glycation increases hemoglobin-oxygen affinity and reduces oxygen delivery to tissues by altering the structure and function of hemoglobin.

Objectives: We investigated whether an elevated blood concentration of glycosylated hemoglobin (HbA1c) could induce falsely high pulse oximeter oxygen saturation (SpO2) in type 2 diabetic patients during mechanical ventilation or oxygen therapy.

Methods: Arterial oxygen saturation (SaO2) and partial pressure of oxygen (PO2) were determined with simultaneous monitoring of SpO2 in 261 type 2 diabetic patients during ventilation or oxygen inhalation.

Results: Blood concentration of HbA1c was >7% in 114 patients and ≤ 7% in 147 patients. Both SaO2 (96.2 ± 2.9%, 95% confidence interval [CI] 95.7-96.7% vs. 95.1 ± 2.8%, 95% CI 94.7-95.6%) and SpO2 (98.0 ± 2.6%, 95% CI 97.6-98.5% vs. 95.3 ± 2.8%, 95% CI 94.9-95.8%) were significantly higher in patients with HbA1c >7% than in those with HbA1c ≤ 7% (Data are mean ± SD, all p < 0.01), but PO2 did not significantly differ between the two groups. Bland-Altman analysis demonstrated a significant bias between SpO2 and SaO2 (1.83 ±0.55%, 95% CI 1.73% -1.94%) and limits of agreement (0.76% and 2.92%) in patients with HbA1c >7%. The differences between SpO2 and SaO2 correlated closely with blood HbA1c levels (Pearson's r = 0.307, p < 0.01).

Conclusions: Elevated blood HbA1c levels lead to an overestimation of SaO2 by SpO2, suggesting that arterial blood gas analysis may be needed for type 2 diabetic patients with poor glycemic control during the treatment of hypoxemia.

Citing Articles

Application of the Single Source-Detector Separation Algorithm in Wearable Neuroimaging Devices: A Step toward Miniaturized Biosensor for Hypoxia Detection.

Nguyen T, Park S, Park J, Sodager A, George T, Gandjbakhche A Bioengineering (Basel). 2024; 11(4).

PMID: 38671806 PMC: 11048477. DOI: 10.3390/bioengineering11040385.


Clinical factors associated with racial differences in the prevalence of occult hypoxemia: a retrospective case-control study.

Mettler S, Charoenngam N, Jaroenlapnopparat A, Tern C, Xanthavanij N, Economidou S medRxiv. 2024; .

PMID: 38585762 PMC: 10996744. DOI: 10.1101/2024.03.28.24305036.


Functional near infrared spectroscopy detects cortical activation changes concurrent with memory loss in postmenopausal women with Type II Diabetes.

Gorniak S, Wagner V, Vaughn K, Perry J, Gulley Cox L, Hibino H Exp Brain Res. 2023; 241(6):1555-1567.

PMID: 37127798 PMC: 10699502. DOI: 10.1007/s00221-023-06581-1.


Overestimation of Oxygen Saturation Measured by Pulse Oximetry in Hypoxemia. Part 1: Effect of Optical Pathlengths-Ratio Increase.

Elron E, Bromiker R, Gleisner O, Yosef-Hai O, Goldberg O, Nitzan I Sensors (Basel). 2023; 23(3).

PMID: 36772474 PMC: 9921559. DOI: 10.3390/s23031434.


Accuracy of Multiple Pulse Oximeters in Stable Critically Ill Patients.

Blanchet M, Mercier G, Delobel A, Nayet E, Bouchard P, Simard S Respir Care. 2023; 68(5):565-574.

PMID: 36596654 PMC: 10171338. DOI: 10.4187/respcare.10582.


References
1.
Kress J, Pohlman A, Hall J . Determination of hemoglobin saturation in patients with acute sickle chest syndrome: a comparison of arterial blood gases and pulse oximetry. Chest. 1999; 115(5):1316-20. DOI: 10.1378/chest.115.5.1316. View

2.
Wilson B, Cowan H, Lord J, Zuege D, Zygun D . The accuracy of pulse oximetry in emergency department patients with severe sepsis and septic shock: a retrospective cohort study. BMC Emerg Med. 2010; 10:9. PMC: 2876142. DOI: 10.1186/1471-227X-10-9. View

3.
BUNN H, GABBAY K, Gallop P . The glycosylation of hemoglobin: relevance to diabetes mellitus. Science. 1978; 200(4337):21-7. DOI: 10.1126/science.635569. View

4.
Bickler P, Feiner J, Severinghaus J . Effects of skin pigmentation on pulse oximeter accuracy at low saturation. Anesthesiology. 2005; 102(4):715-9. DOI: 10.1097/00000542-200504000-00004. View

5.
Bland J, Altman D . Measuring agreement in method comparison studies. Stat Methods Med Res. 1999; 8(2):135-60. DOI: 10.1177/096228029900800204. View