Dark Skin Decreases the Accuracy of Pulse Oximeters at Low Oxygen Saturation: the Effects of Oximeter Probe Type and Gender

Overview
Journal Anesth Analg
Specialty Anesthesiology
Date 2007 Dec 6
PMID 18048893
Citations 140
Authors
Affiliations
Soon will be listed here.
Abstract

Introduction: Pulse oximetry may overestimate arterial oxyhemoglobin saturation (Sao2) at low Sao2 levels in individuals with darkly pigmented skin, but other factors, such as gender and oximeter probe type, remain less studied.

Methods: We studied the relationship between skin pigment and oximeter accuracy in 36 subjects (19 males, 17 females) of a range of skin tones. Clip-on type sensors and adhesive/disposable finger probes for the Masimo Radical, Nellcor N-595, and Nonin 9700 were studied. Semisupine subjects breathed air-nitrogen-CO2 mixtures via a mouthpiece to rapidly achieve 2- to 3-min stable plateaus of Sao2. Comparisons of Sao2 measured by pulse oximetry (Spo2) with Sao2 (by Radiometer OSM-3) were used in a multivariate model to assess the source of errors.

Results: The mean bias (Spo2 - Sao2) for the 70%-80% saturation range was 2.61% for the Masimo Radical with clip-on sensor, -1.58% for the Radical with disposable sensor, 2.59% for the Nellcor clip, 3.6% for the Nellcor disposable, -0.60% for the Nonin clip, and 2.43% for the Nonin disposable. Dark skin increased bias at low Sao2; greater bias was seen with adhesive/disposable sensors than with the clip-on types. Up to 10% differences in saturation estimates were found among different instruments in dark-skinned subjects at low Sao2.

Conclusions: Multivariate analysis indicated that Sao2 level, sensor type, skin color, and gender were predictive of errors in Spo2 estimates at low Sao2 levels. The data suggest that clinically important bias should be considered when monitoring patients with saturations below 80%, especially those with darkly pigmented skin; but further study is needed to confirm these observations in the relevant populations.

Citing Articles

Monte Carlo Simulation of the Effect of Melanin Concentration on Light-Tissue Interactions in Reflectance Pulse Oximetry.

Al-Halawani R, Qassem M, Kyriacou P Sensors (Basel). 2025; 25(2).

PMID: 39860931 PMC: 11769186. DOI: 10.3390/s25020559.


Leveraging Smart Telemedicine Technology to Enhance Nursing Care Satisfaction and Revolutionize COVID-19 Care: Prospective Cohort Study.

Chang Y, Lin C, Hsu J, Liao S, Yu C, Peng H JMIR Hum Factors. 2025; 12:e53456.

PMID: 39838732 PMC: 11774322. DOI: 10.2196/53456.


Accuracy of Rhythm Diagnostic Systems' MultiSense in Detection of Arterial Oxygen Saturation and Respiratory Rate During Hypoxia in Humans: Effects of Skin Color and Device Localization.

Evrard C, El Attaoui A, Pistea C, Enache I, Marriott M, Mayaud L Sensors (Basel). 2025; 25(1.

PMID: 39796917 PMC: 11722614. DOI: 10.3390/s25010127.


Impact of Skin Pigmentation on Pulse Oximetry Blood Oxygenation and Wearable Pulse Rate Accuracy: Systematic Review and Meta-Analysis.

Singh S, Bennett M, Chen C, Shin S, Ghanbari H, Nelson B J Med Internet Res. 2024; 26:e62769.

PMID: 39388258 PMC: 11502980. DOI: 10.2196/62769.


Exploring the impact and influence of melanin on frequency-domain near-infrared spectroscopy measurements.

Roy S, Wu J, Cao J, Disu J, Bharadwaj S, Meinert-Spyker E J Biomed Opt. 2024; 29(Suppl 3):S33310.

PMID: 39323492 PMC: 11423252. DOI: 10.1117/1.JBO.29.S3.S33310.