The Adenosine Triphosphate Test is a Rapid and Reliable Audit Tool to Assess Manual Cleaning Adequacy of Flexible Endoscope Channels

Overview
Date 2012 Sep 15
PMID 22975364
Citations 23
Authors
Affiliations
Soon will be listed here.
Abstract

Background: The study objective was to verify that the adenosine triphosphate (ATP) benchmark of <200 relative light units (RLUs) was achievable in a busy endoscopy clinic that followed the manufacturer's manual cleaning instructions.

Methods: All channels from patient-used colonoscopes (20) and duodenoscopes (20) in a tertiary care hospital endoscopy clinic were sampled after manual cleaning and tested for residual ATP. The ATP test benchmark for adequate manual cleaning was set at <200 RLUs. The benchmark for protein was <6.4 μg/cm(2), and, for bioburden, it was <4-log10 colony-forming units/cm(2).

Results: Our data demonstrated that 96% (115/120) of channels from 20 colonoscopes and 20 duodenoscopes evaluated met the ATP benchmark of <200 RLUs. The 5 channels that exceeded 200 RLUs were all elevator guide-wire channels. All 120 of the manually cleaned endoscopes tested had protein and bioburden levels that were compliant with accepted benchmarks for manual cleaning for suction-biopsy, air-water, and auxiliary water channels.

Conclusion: Our data confirmed that, by following the endoscope manufacturer's manual cleaning recommendations, 96% of channels in gastrointestinal endoscopes would have <200 RLUs for the ATP test kit evaluated and would meet the accepted clean benchmarks for protein and bioburden.

Citing Articles

A comparative assessment of contamination rates in gastrointestinal endoscope reprocessing: sterilization versus high-level disinfection.

Chantarojanasiri T, Rungrueangmaitree R, Thongsri S, Jampa-Ngern U, Ratanachu-Ek T DEN Open. 2025; 5(1):e70093.

PMID: 40066176 PMC: 11891260. DOI: 10.1002/deo2.70093.


Beyond Endoscopes: Pilot Study of Surgical Instrument Lumen Inspection.

Hopkins K, Adams S, Lamb L, Smart A, Ofstead C Biomed Instrum Technol. 2024; 58(1):25-33.

PMID: 38377297 PMC: 10878677. DOI: 10.2345/0899-8205-58.1.25.


Assessment of endoscope reprocessing at World Gastroenterology Organisation training centers using adenosine triphosphate testing.

Visrodia K, Jones D, Jennings M, Conway C, Burgos H, Malani J Endosc Int Open. 2024; 12(1):E68-E77.

PMID: 38193007 PMC: 10774016. DOI: 10.1055/a-2185-6035.


Effect of a disposable endoscope precleaning kit in the cleaning procedure of gastrointestinal endoscope: A multi-center observational study.

Wang Y, Wu Y, Liu X, Li J, Zhan Y, Liu B World J Gastrointest Endosc. 2024; 15(12):705-714.

PMID: 38187912 PMC: 10768042. DOI: 10.4253/wjge.v15.i12.705.


Adenosine triphosphate (ATP) sampling algorithm for monitoring the cleanliness of surgical instruments.

Pontes D, Costa D, Pereira P, Whiteley G, Glasbey T, Tipple A PLoS One. 2023; 18(8):e0284967.

PMID: 37582099 PMC: 10426997. DOI: 10.1371/journal.pone.0284967.