» Articles » PMID: 22915946

Validation of the BPLab(®) 24-hour Blood Pressure Monitoring System According to the European Standard BS EN 1060-4:2004 and British Hypertension Society Protocol

Overview
Publisher Dove Medical Press
Date 2012 Aug 24
PMID 22915946
Citations 13
Authors
Affiliations
Soon will be listed here.
Abstract

Introduction: Automatic blood pressure (BP) measuring devices are more and more often used in BP self-checks and in 24-hour BP monitoring. Nowadays, 24-hour BP monitoring is a necessary procedure in arterial hypertension treatment. The aim of this study was to validate the BPLab(®) ambulatory blood pressure monitor according to the European standard BS EN 1060-4:2004 and the British Hypertension Society (BHS) protocol, as well as to work out solutions regarding the suitability of using this device in clinical practice.

Methods: A group of 85 patients of both sexes and different ages, who voluntarily agreed to take part in the tests and were given detailed instructions on the measurement technique were recruited for this study. The results of the BP measurement obtained by a qualified operator using the BPLab(®) device were compared with the BP values measured using the Korotkov auscultatory method. Data were obtained simultaneously by two experts with experience of over 10 years and had completed a noninvasive BP measurement standardization training course. Discrepancies in the systolic and diastolic BP measurements (N = 510; 255 for each expert) were analyzed according to the criteria specified in the BHS-93 protocol.

Results: The device passed the requirements of the European Standard BS EN 1060-4:2004 and was graded 'A' according to the criteria of the BHS protocol for both systolic BP and diastolic BP.

Conclusion: The BPLab(®) 24-hour ambulatory blood pressure monitoring device may be recommended for extensive clinical use.

Citing Articles

Relationship of choroidal thickness with pulsatile hemodynamics in essential hypertensive patients.

Mule G, Vadala M, Sinatra N, Mancia E, Sorce A, Geraci G J Clin Hypertens (Greenwich). 2021; 23(5):1030-1038.

PMID: 33492773 PMC: 8678803. DOI: 10.1111/jch.14196.


Ambulatory blood pressure and arterial stiffness web-based telemonitoring in patients at cardiovascular risk. First results of the VASOTENS (Vascular health ASsessment Of The hypertENSive patients) Registry.

Omboni S, Posokhov I, Parati G, Rogoza A, Kotovskaya Y, Arystan A J Clin Hypertens (Greenwich). 2019; 21(8):1155-1168.

PMID: 31294910 PMC: 8030443. DOI: 10.1111/jch.13623.


Relationships between 24-h blood pressure variability and 24-h central arterial pressure, pulse wave velocity and augmentation index in hypertensive patients.

Omboni S, Posokhov I, Rogoza A Hypertens Res. 2016; 40(4):385-391.

PMID: 27881851 DOI: 10.1038/hr.2016.156.


Twenty-Four-Hour Ambulatory Pulse Wave Analysis in Hypertension Management: Current Evidence and Perspectives.

Omboni S, Posokhov I, Kotovskaya Y, Protogerou A, Blacher J Curr Hypertens Rep. 2016; 18(10):72.

PMID: 27659178 DOI: 10.1007/s11906-016-0681-2.


Vascular Health Assessment of The Hypertensive Patients (VASOTENS) Registry: Study Protocol of an International, Web-Based Telemonitoring Registry for Ambulatory Blood Pressure and Arterial Stiffness.

Omboni S, Posokhov I, Parati G, Avolio A, Rogoza A, Kotovskaya Y JMIR Res Protoc. 2016; 5(2):e137.

PMID: 27358088 PMC: 4945820. DOI: 10.2196/resprot.5619.


References
1.
OBrien E, Mee F, Tan K, Atkins N, OMalley K . Training and assessment of observers for blood pressure measurement in hypertension research. J Hum Hypertens. 1991; 5(1):7-10. View

2.
Pickering T, Miller N, Ogedegbe G, Krakoff L, Artinian N, Goff D . Call to action on use and reimbursement for home blood pressure monitoring: a joint scientific statement from the American Heart Association, American Society of Hypertension, and Preventive Cardiovascular Nurses Association. J Cardiovasc Nurs. 2008; 23(4):299-323. DOI: 10.1097/01.JCN.0000317429.98844.04. View

3.
OBrien E . Replacing the mercury sphygmomanometer. Requires clinicians to demand better automated devices. BMJ. 2000; 320(7238):815-6. PMC: 1127179. DOI: 10.1136/bmj.320.7238.815. View

4.
OBrien E, Petrie J, Littler W, de Swiet M, Padfield P, OMalley K . The British Hypertension Society protocol for the evaluation of automated and semi-automated blood pressure measuring devices with special reference to ambulatory systems. J Hypertens. 1990; 8(7):607-19. DOI: 10.1097/00004872-199007000-00004. View

5.
OBrien E, Petrie J, Littler W, de Swiet M, Padfield P, Altman D . An outline of the revised British Hypertension Society protocol for the evaluation of blood pressure measuring devices. J Hypertens. 1993; 11(6):677-9. DOI: 10.1097/00004872-199306000-00013. View