Monte Carlo Study of In-field and Out-of-field Dose Distributions from a Linear Accelerator Operating with and Without a Flattening-filter

Overview
Journal Med Phys
Specialty Biophysics
Date 2012 Aug 17
PMID 22894444
Citations 13
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: To compare dosimetric characteristics of 6 MV photon fields originating from a linear accelerator operating with (FF) and without (FFF) a flattening-filter. The main objective is to establish a FFF model that results in similar depth-dose and build-up profiles as the original FF model, and subsequently estimate and compare out-of-field dose distributions.

Methods: The EGSnrc Monte Carlo user codes BEAMnrc and DOSXYZnrc are used for photon beam simulations of an Elekta linear accelerator and dose calculations in a water phantom, respectively. Three beam models were analyzed: (1) the conventional linear accelerator with the flattening-filter in place and incident electron energy 6.45 MeV (FF 6.45 MeV), (2) similar flattening-filter-free model (FFF 6.45 MeV), and (3) as (2) but with increased electron energy (FFF 8.0 MeV). The field size 5 × 5 cm(2) was used for characterization of dose output, depth dose profiles, and photon spectrum. The field size 40 × 40 cm(2) was used for characterization of cross-field photon energy, photon fluence, and dose distributions. Out-of-field dose distributions were analyzed in both in-plane and cross-plane directions for 5 × 5 cm(2) and 10 × 10 cm(2) fields.

Results: Comparable depth dose distributions, including the build-up region, for FF and FFF fields were achieved by increasing the electron energy from 6.45 MeV to 8.0 MeV for the FFF beam. The FFF beams result in reduced out-of-field dose compared to the FF beam: the reduction was most apparent in the cross-plane direction and more pronounced by the FFF 8.0 MeV beam compared to the FFF 6.45 MeV beam. Differences in out-of-field dose due to direction (in-plane vs cross-plane) were up to 40% for the FF beam; this effect was significantly reduced for the FFF beams. As the flattening-filter is a major source of contaminating electrons, superficial out-of-field dose was expected, and was found to be, reduced for FFF beams.

Conclusions: The build-up and depth-dose characteristics of a conventional "6 MV" beam can be maintained when changing to a flattening-filter-free modality by increasing the incident electron energy from 6.45 MeV to 8.0 MeV. This will at the same time reduce the out-of-field dose for regions up to 20 cm from the central axis by 10%-30% compared to the original FF situation.

Citing Articles

Dosimetric characteristics of 6MV flattening filter free and flattened beams among beam-matched linacs: a three-institutional study.

Ghemis D, Marcu L, Virag V, Virag A Radiat Oncol. 2023; 18(1):126.

PMID: 37507741 PMC: 10375603. DOI: 10.1186/s13014-023-02313-5.


Sensitivity Analysis of a 6 MeV Photon Beam Monte Carlo Model.

Shami N, Atarod M, Shokrani P, Najafizade N J Med Signals Sens. 2023; 13(2):144-152.

PMID: 37448550 PMC: 10336916. DOI: 10.4103/jmss.jmss_152_21.


Analytical models for external photon beam radiotherapy out-of-field dose calculation: a scoping review.

Benzazon N, Colnot J, de Kermenguy F, Achkar S, De Vathaire F, Deutsch E Front Oncol. 2023; 13:1197079.

PMID: 37228501 PMC: 10203488. DOI: 10.3389/fonc.2023.1197079.


Benchmarking of electron beam parameters based on Monte Carlo linear accelerator simulation.

Zhang F, Zhou M, Liu J, Yue L, Deng L, Xu Z Transl Cancer Res. 2022; 9(2):577-584.

PMID: 35117402 PMC: 8798818. DOI: 10.21037/tcr.2019.12.02.


Comparison of Flattening Filter and Flattening Filter-Free Volumetric Modulated Arc Radiotherapy in Patients with Locally Advanced Nasopharyngeal Carcinoma.

Jia F, Xu D, Yue H, Wu H, Li G Med Sci Monit. 2018; 24:8500-8505.

PMID: 30472719 PMC: 6276721. DOI: 10.12659/MSM.910218.