Absorption-mode Spectra on the Dynamically Harmonized Fourier Transform Ion Cyclotron Resonance Cell

Overview
Specialty Chemistry
Date 2012 Aug 1
PMID 22847701
Citations 10
Authors
Affiliations
Soon will be listed here.
Abstract

Rationale: The recently designed dynamically harmonized Fourier transform ion cyclotron resonance (FT-ICR) cell creates a more harmonized electric field for the detection of the cyclotron motion of ions and prolongs the ion transient from seconds to minutes. In order to achieve its best performance, phase correction was applied in the spectra, and new advantages of the absorption-mode were revealed.

Methods: Spectra were acquired from both simple standard and complex mixtures using either narrowband or broadband mode, and the data were processed to compare the performance of the spectra in magnitude and absorption-mode.

Results: The research shows that phase correction works well with data from Nikolaev's new cell, which produces the maximum improvement in resolving power (2×), and improves the match with the theoretical intensities of the isotopic peaks. In addition, the harmonic peaks can be diagnosed immediately in the absorption-mode.

Conclusions: The manuscript demonstrates absorption-mode spectra from Nikolaev's ICR cell, which will be of interest to the community. The improved relative peak intensities and immediate identification of harmonic peaks will facilitate data interpretation.

Citing Articles

Fellgett Revisited: On the Nature of Noise in Two-Dimensional Mass Spectrometry.

Littlejohn C, Li M, Lam P, Barrow M, OConnor P J Am Soc Mass Spectrom. 2024; 35(12):2984-2992.

PMID: 39454130 PMC: 11622379. DOI: 10.1021/jasms.4c00294.


Quadrupole detection FT-ICR mass spectrometry offers deep profiling of residue oil: A systematic comparison of 2ω 7 Tesla versus 15 Tesla instruments.

Ge J, Ma C, Qi Y, Wang X, Wang W, Hu M Anal Sci Adv. 2024; 2(5-6):272-278.

PMID: 38716153 PMC: 10989625. DOI: 10.1002/ansa.202000123.


Ultra-Accurate Correlation between Precursor and Fragment Ions in Two-Dimensional Mass Spectrometry: Acetylated vs Trimethylated Histone Peptides.

Palasser M, Heel S, Delsuc M, Breuker K, van Agthoven M J Am Soc Mass Spectrom. 2023; 34(4):608-616.

PMID: 36930827 PMC: 10080674. DOI: 10.1021/jasms.2c00319.


Fine Structure in Isotopic Peak Distributions Measured Using Fourier Transform Ion Cyclotron Resonance Mass Spectrometry: A Comparison between an Infinity ICR Cell and a Dynamically Harmonized ICR Cell.

Xu J, Li M, Marzullo B, Wootton C, Barrow M, OConnor P J Am Soc Mass Spectrom. 2022; 33(8):1499-1509.

PMID: 35763614 PMC: 9354249. DOI: 10.1021/jasms.2c00093.


Phase Correction for Absorption Mode Two-Dimensional Mass Spectrometry.

Delsuc M, Breuker K, van Agthoven M Molecules. 2021; 26(11).

PMID: 34205070 PMC: 8199897. DOI: 10.3390/molecules26113388.