Computational Fluid Dynamics Analysis to Determine Shear Stresses and Rates in a Centrifugal Left Ventricular Assist Device

Overview
Journal Artif Organs
Date 2012 Feb 25
PMID 22360826
Citations 11
Authors
Affiliations
Soon will be listed here.
Abstract

Axial flow left ventricular assist devices (LVADs) are a significant improvement in mechanical circulatory support. However, patients with these devices experience degradation of large von Willebrand factor (vWF) multimers, which is associated with bleeding and may be caused by high shear stresses within the LVAD. In this study, we used computational fluid mechanics to determine the wall shear stresses, shear rates, and residence times in a centrifugal LVAD and assess the impact on these variables caused by changing impeller speed and changing from a shrouded to a semi-open impeller. In both LVAD types, shear rates were well over 10,000/s in several regions. This is high enough to degrade vWF, but it is unclear if residence times, which were below 5ms in high-shear regions, are long enough to allow vWF cleavage. Additionally, wall shear stresses were below the threshold stress of 10Pa only in the outlet tube so it is feasible to endothelialize this region to enhance its biocompatibility.

Citing Articles

The dynamics of red blood cells traversing slits of mechanical heart valves under high shear.

Meng K, Chen H, Pan Y, Li Y Biophys J. 2024; 123(21):3780-3797.

PMID: 39340153 PMC: 11560308. DOI: 10.1016/j.bpj.2024.09.027.


Physical Properties of Blood and their Relationship to Clinical Conditions.

Alexy T, Detterich J, Connes P, Toth K, Nader E, Kenyeres P Front Physiol. 2022; 13:906768.

PMID: 35874542 PMC: 9298661. DOI: 10.3389/fphys.2022.906768.


A Novel Fragmentation Sensitivity Index Determines the Susceptibility of Red Blood Cells to Mechanical Trauma.

Ugurel E, Goksel E, Goktas P, Cilek N, Atar D, Yalcin O Front Physiol. 2021; 12:714157.

PMID: 34512387 PMC: 8424113. DOI: 10.3389/fphys.2021.714157.


Hemolysis induced by Left Ventricular Assist Device is associated with proximal tubulopathy.

de Nattes T, Litzler P, Gay A, Nafeh-Bizet C, Francois A, Guerrot D PLoS One. 2020; 15(11):e0242931.

PMID: 33253314 PMC: 7703997. DOI: 10.1371/journal.pone.0242931.


Computational modeling of the Food and Drug Administration's benchmark centrifugal blood pump.

Good B, Manning K Artif Organs. 2020; 44(7):E263-E276.

PMID: 31971269 PMC: 7308203. DOI: 10.1111/aor.13643.


References
1.
Burgreen G, Loree 2nd H, Bourque K, Dague C, Poirier V, Farrar D . Computational fluid dynamics analysis of a maglev centrifugal left ventricular assist device. Artif Organs. 2004; 28(10):874-80. DOI: 10.1111/j.1525-1594.2004.07384.x. View

2.
Siedlecki C, Lestini B, Eppell S, Wilson D, Marchant R . Shear-dependent changes in the three-dimensional structure of human von Willebrand factor. Blood. 1996; 88(8):2939-50. View

3.
Bourque K, Gernes D, Loree H, Richardson J, Poirier V, Barletta N . Incorporation of electronics within a compact, fully implanted left ventricular assist device. Artif Organs. 2002; 26(11):939-42. DOI: 10.1046/j.1525-1594.2002.07118.x. View

4.
Vincentelli A, Susen S, Le Tourneau T, Six I, Fabre O, Juthier F . Acquired von Willebrand syndrome in aortic stenosis. N Engl J Med. 2003; 349(4):343-9. DOI: 10.1056/NEJMoa022831. View

5.
Curtas A, Wood H, Allaire P, McDaniel J, Day S, Olsen D . Computational fluid dynamics modeling of impeller designs for the HeartQuest left ventricular assist device. ASAIO J. 2002; 48(5):552-61. DOI: 10.1097/00002480-200209000-00019. View