Comparison of Methods: Passing and Bablok Regression

Overview
Specialty Biochemistry
Date 2011 Dec 7
PMID 22141206
Citations 108
Authors
Affiliations
Soon will be listed here.
Abstract

The comparison of methods experiment is important part in process of analytical methods and instruments validation. Passing and Bablok regression analysis is a statistical procedure that allows valuable estimation of analytical methods agreement and possible systematic bias between them. It is robust, non-parametric, non sensitive to distribution of errors and data outliers. Assumptions for proper application of Passing and Bablok regression are continuously distributed data and linear relationship between data measured by two analytical methods. Results are presented with scatter diagram and regression line, and regression equation where intercept represents constant and slope proportional measurement error. Confidence intervals of 95% of intercept and slope explain if their value differ from value zero (intercept) and value one (slope) only by chance, allowing conclusion of method agreement and correction action if necessary. Residual plot revealed outliers and identify possible non-linearity. Furthermore, cumulative sum linearity test is performed to investigate possible significant deviation from linearity between two sets of data. Non linear samples are not suitable for concluding on method agreement.

Citing Articles

Therapeutic drug monitoring and immunogenetic factors associated with the use of adalimumab in Crohn's disease patients.

Genaro L, Carron J, de Castro M, Franceschini A, Lourenco G, Cruz C Int J Immunopathol Pharmacol. 2025; 39:3946320251319379.

PMID: 39959979 PMC: 11831650. DOI: 10.1177/03946320251319379.


Comparison of 3 methods of quantification of phenobarbital in canine plasma: high-performance liquid chromatography, a point-of-care immunoassay, and the FDA-approved immunoassay analyzer.

Thungrat K, Gibson R, Jukier T, Jenkins A, Cruz-Espindola C J Vet Diagn Invest. 2025; 37(2):272-277.

PMID: 39866002 PMC: 11773498. DOI: 10.1177/10406387241312891.


Comparison of Freestyle Optium Neo H and Centrivet GK Device in the Diagnosis of Hypoglycaemia and Hyperketonaemia in Dairy Goats: A Field Study.

Uztimur M, Unal C, Deniz A, Pekmezci A Vet Med Sci. 2025; 11(1):e70159.

PMID: 39792046 PMC: 11720734. DOI: 10.1002/vms3.70159.


Inductively Coupled Plasma Mass Spectrometry Performance for the Measurement of Key Serum Minerals: A Comparative Study With Standard Quantification Methods.

Kojo K, Oguri T, Tanaka T, Ikeda A, Shimizu T, Fujimoto S J Clin Lab Anal. 2024; 39(2):e25140.

PMID: 39716823 PMC: 11776499. DOI: 10.1002/jcla.25140.


Analytical and clinical performance of eight Simoa and Lumipulse assays for automated measurement of plasma p-tau181 and p-tau217.

Wojdala A, Vanbrabant J, Bayoumy S, Antwi-Berko D, Le Bastard N, van der Flier W Alzheimers Res Ther. 2024; 16(1):266.

PMID: 39702308 PMC: 11660642. DOI: 10.1186/s13195-024-01630-5.